Connecting humidity sensor to 10 channels
QIs it possible to measure 10 channels of humidity by combining Memory HiLogger LR8450 and Universal Unit U8551 and connecting ten Humidity Sensor Z2000? How much current can be supplied from the LR8450 to the Humidity Sensor Z2000?
A
Yes, you can.
Since the current consumption of the Z2000 is less than 1 mA, it can supply power for 100 units from the voltage output terminal of the LR8450.
The maximum supply current of the LR8450 is 100 mA.
However, since the voltage output terminal is a push-button type, it is necessary to solder multiple wires together.