LCR METER IM3523A
V1.04 May 22, 2026
- If you are unable to download software, please check the following:
- Please register a my HIOKI account and then use it to log in to the my HIOKI service.
| Firmware (IM3523AV104.zip) | |
|---|---|
| Version Upgrade Procedure (IM3523A_Vup_en_01.pdf) |
・All software provided here is freeware and the best version currently available.
・If the problems or errors during this download and install lead to malfunction, please contact your dealer or Hioki representative.
・The ownership, intellectual right and all other associated rights relating to the software provided here belong to Hioki E.E. Corporation.
If you have read and understand the information above and wish to proceed then please download the designated files.
As for version upgrade procedure, please download "Procedure on how to update the firmware" which describes the preparation and procedure for upgrading firmware after downloading the files.
Precaution when updating firmware:
・To conduct the version upgrade, USB cable (Type A/B) is required.
・During the version upgrade process, don't turn off the power to the IM3523A or don't remove the USB cable from the connector.
Firmware for the Model IM3523A
The latest firmware version is shown below. A version update is highly recommended if you are running older version.
・Firmware Version : V1.04
How to check the version number
The version number of firmware can be checked by selecting the ”INFO” tab of the System screen.
Version Upgrade History
V1.04 (May 2026)
Bug Fixes
1. Fixed a bug where the system would hang if the “Panel Load” command was repeated via communication.
2. Fixed a bug where executing the “Panel Save” command via communication after calibration would save the data from the calibration process.
3. Fixed a bug where calibration values were not applied to the first measurement when loading a panel containing only ADJ via EXT. I/O.
4. Fixed a bug where measurements might not be performed correctly under the second set of measurement conditions during continuous measurement.
Functional Improvements
1. Improved the display of numbers during panel loading.
2. Reduced panel loading time.
V1.03 (October 2025)
1. Fixed a bug where the lighting status of the link LED on the LAN connector could sometimes reverse.
2. Added a setting to disable auto-negotiation for LAN communication, allowing for communication connections using 100BASE-TX full duplex with fixed MDI.
V1.02 (December 2023)
1. Fixed a bug where unnecessary spaces were included in the command to respond to measurements.
2. When the IP address is changed, a packet (GARP) notifying the update is now output.
3. GENNECT ONE (V5.30 or later) is now supported.
V1.01 (January 2023)
1.Release