RESISTANCE HiTESTER RM3542
Measure in as Fast as 0.9 ms, Optimized for Automated Systems
Hioki DC resistance meters measure a broad range of resistance values at a high level of precision. The RM3542 deliver high speed, high precision, and high reliability, indispensable for the shipping inspection of devices such as resistors and EMI components. The resistance meter also supports chip resistor sampling inspections, helping resistor manufacturers improve productivity.
Key Features
- High speed and accuracy maximize productivity in automated systems
- Multiple checking functions ensure proper contact for reliable measurements
- Low-power resistance mode measures chip inductors and EMC suppression components
- Supports sample inspections during the manufacturing process
Model No. (Order Code)
RM3542 | |
---|---|
RM3542-01 | Built in GP-IB interface |
Ensure reliable contact quickly and reduce the contact error rate with the contact improvement function
The RM3542 improves contact with measurement specimens by eliminating oxide films and dirt between the probes and specimens. By addressing faulty contact, the instrument facilitates stable measurement and reduces the contact error rate, boosting productivity.
Reduce Measurement Time for More Stable Measurements with Integration Time Setting Function
Set the integration time as desired for each range - SHORT to reduce the takt time, or LONG for more stable inspections.
Basic specifications
Resistance range | [at Low Power OFF] 100 mΩ range (Max. 120.0000 mΩ, 0.1 μΩ resolution) to 100 MΩ range (Max. 120.0000 MΩ, 100 Ω resolution), 10 steps [at Low Power ON] 1000 mΩ range (Max. 1200.000 mΩ, 1 μΩ resolution) to 1000 Ω range (Max. 1200.000 Ω, 1 mΩ resolution), 4 steps |
|||||
---|---|---|---|---|---|---|
Display | Monochrome graphic LCD 240 × 64 dot, white LED backlight | |||||
Measurement accuracy | [with SLOW mode, at 100 mΩ range] ±0.015 % rdg. ±0.002 % f.s. [with SLOW mode, at 1000 Ω range] ±0.006 % rdg. ±0.001 % f.s. (the best case) |
|||||
Testing current | [at 100 mΩ range] 100 mA DC to [at 100 MΩ range] 100 nA DC | |||||
Open-terminal voltage | 20 V DC max. | |||||
Sampling rate | FAST, MEDIUM, SLOW, 3 settings | |||||
Measurement times | [at 100 Ω /1000 Ω ranges, with Low Power OFF] FAST: 0.9 ms, MED: 3.6 ms, SLOW: 17 ms (minimum time) |
|||||
Integration time | 0.1 ms to 100.0 ms, or 1 to 5 PLC at 50 Hz, 1 to 6 PLC at 60 Hz Note: PLC = one power line cycle (mains wave-form period) |
|||||
Other functions | Comparator (compare setting value with measurement value), Delay (set to allow for mechanical delay of trigger input and probing, or set to allow for measurement object response), OVC (offset voltage compensation), Measurement fault detection, Probe short-circuit detection, Improve contact, Auto-memory, Statistical calculations, Settings monitor (when using two instruments, a difference in settings causes warning notification), Retry, Trigger function. etc,. | |||||
Interfaces | RS-232C, Printer, GP-IB (Model RM3542-01) | |||||
External I/O | Trigger, Hold input, Comparator output, Settings monitor terminal | |||||
Power supply | 100 to 240 V AC, 50/60 Hz, 30 VA max. | |||||
Dimensions and mass | 260 mm (10.24 in) W × 88 mm (3.46 in) H × 300 mm (11.81 in) D, 2.9 kg (102.3 oz) | |||||
Included accessories | Power cord ×1, EXT. I/O male connector ×1, Instruction manual ×1, Operation guide ×1 |
Probe and Test fixtures (3)
DC to 100 kHz, 1 m (3.28 ft) length, impedance characteristics of 75 Ω
Direct connection type, DC to 8 MHz(Can be measured up to 10 MHz, when combined with the special order model IM3536-01), measurable conductor diameter: ø0.3 (0.01 in) to 2 mm (0.08 in)
Direct connection type, DC to 8 MHz(Can be measured up to 10 MHz, when combined with the special order model IM3536-01), Test sample dimensions:1 mm (0.04 in) to 10 mm (0.39 in)
PC communication (2)
2 m (6.56 ft) length
9 pin - 9 pin, cross, 1.8 m (5.91 ft) length