Evaluation of Input and Output Characteristics of Power Conditioners
One power analyzer can perform completely synchronized measurement of input and output characteristics of power conditioners.
■Highlights
- The 3390 Power Analyzer can simultaneously display up to 32 items, including DC/AC voltage/current,
voltage ripple factor, power, amount of power energy, power factor, voltage/current distortion factors,
frequency, amount of power energy sold and purchased, voltage imbalance rate, and efficiency.
- Also display voltage/current waveforms.
■Connection example
■How to use
- Select the appropriate current sensor according to the type and level of electrical current. 
 - AC/DC Clamp-on current sensor : 9277(20A), 9278(200A), 9279(500A)
 - AC/DC Feed-through current sensor : CT6862(50A), CT6863(200A), 9709(500A), CT6865(1000A) 
 - AC Clamp-on current sensor : 9272-10(20/200A) 
- Set the integration mode according to the AC and DC. 
This setting automatically changes the display of the
ripple factor and total harmonic distortion factor. 
 - DC mode : Ripple factor
 (input and internal measurement channel) 
 - RMS mode: Total harmonic distortion factor
 (output measurement channel) 
- Configure the settings for efficiency measurement
 and loss measurement of the power conditioner. 
 - Pin : Power on the input measurement channel 
 - Pout: Power on the output measurement channel
      
                   
      
 
          

 
       
       
       
       
       
       
       
       
       
       
       
      