BATTERY HiTESTER 3555
Portable Battery Tester for Testing Compact Storage Batteries for Mobile Devices
Discontinued
Hioki portable battery testers support the maintenance of UPS and storage batteries critical to the life support of all businesses. The 3551 is a 300 mΩ to 30 Ω/30V portable battery internal resistance tester suitable for testing high-capacity alkali and lead-acid storage batteries used in mobile devices.
Key Features
- For use with compact batteries such as NiCad and nickel-metal hydride
- Instantaneously diagnose degradation (PASS, CAUTION, FAIL) by measuring internal resistance and voltage (*1)
- *1:Operator must input the criteria for PASS/FAIL judgments according to the type of battery being measured.
Model No. (Order Code)
3555 | Discontinued |
---|
Basic specifications
Resistance measurement range | 300 mΩ (max. display 300.00 mΩ, resolution 100 μΩ) to 30 Ω (max. display 30.00 Ω, resolution 10 mΩ), 3 ranges Accuracy: ±1.2 % rdg. ±6 dgt. (All ranges) Testing source frequency: 1 kHz ±5 Hz, testing current: 5 mA (300 mΩ range), 500 μA (3 Ω range), 50 μA (30 Ω range) Open-circut Voltage: 5 V peak |
|||||
---|---|---|---|---|---|---|
Voltage measurement range | ± 3 V (max. display ±3.000 V, resolution: 1 mV) to ± 30 V (max. display ±30.00 V, resolution: 10 mV), 2 ranges Accuracy: ±0.15 % rdg. ±6 dgt. |
|||||
Absolute max. input voltage | 50 V DC max. (No AC input) | |||||
Sampling rate | 1.25 times/s (at resistance and voltage measurement combination) | |||||
Comparator functions | Setting: Upper and lower limit for resistance, and lower limit for voltage, Number of comparator setting: 10 sets Comparator output: Pass (green), Warning (amber), and Fail (red) LEDs, Audible output on warning or Fail |
|||||
Other | Data storage: N/A, Interface: N/A, Temperature measurement: N/A | |||||
Power supply | LR6 (AA) Alkaline dry battery × 6, Continuous operating time: 18 h | |||||
Dimensions and mass | 196 mm (7.72 in) W × 130 mm (5.12 in) H × 50 mm (1.97 in) D, 680 g (24.0 oz) (including batteries) | |||||
Included accessories | Pin Type Lead 9461 × 1, Instruction manual × 1, LR6 (AA) alkaline battery × 6 |
Measurement Leads (4)
*The L2107 cannot use with Model 3555
About probe length A: From junction to probe B: Probe part L: Whole length

A:220 mm (8.66 in), B:197 mm (7.76 in), L:1360 mm (4.46 ft)

A: 280 mm (11.02 in.), B: 118 mm (4.65 in.), L: 1360 mm (4.46 ft.), 60 V DC

For the L2100, 9465-10, 9465, 9461

A:240 mm (9.45 in), B:132 mm (5.20 in), L:804 mm (2.64 ft)
Primarily for use with compact secondary batteries, etc. (2)
(for terminals with very little surface area)
About probe length A: From junction to probe B: Probe part L: Whole length

A: 260 mm (10.24 in.), B: 140 mm (5.51 in.), L: 850 mm (2.79 ft.), 60 V DC