Resistance Measurement Range Settings, Communication Commands

QI’m communicating the Resistance HiTester RM3543 with the RS-232C interface, but it does not work even if I set the command (message) of "Measurement range". What are the possible causes?

A

Are you describing [Measurement Range] in the command setting of the resistance measurement range of the Resistance HiTester RM3543? In that case, please describe the [Expected measurement value] (resistance value) instead of the [Measurement Range]. This sets the RM3543 main instrument to the optimum measuring range.

Note: The above applies equally to other resistance meters such as RM3545 and RM3542. Moreover, the LCR Meter inputs the impedance measurement range.