RM3544 Firmware Update

V1.04, V2.02 Jun 10, 2022

If you are unable to download software, please check the following:
  • Please register a my HIOKI account and then use it to log in to the my HIOKI service.
  • Product firmware (product upgrades) and computer application software can only be downloaded for products you have registered on the my HIOKI service.
Firmware (RM3544V104_V202.ZIP)
Version update procedure (RM3544Vup_en.pdf)

○ Latest software is posted. Keep the instruction manual up-to-date.
○ Do not perform the following operations during the upgrade. Repair is required.
・ Turn off the main unit.
・ Unplug the USB cable.

Specifications

・Upgrade version V1.02, V1.03 to V1.04.
Or upgrade the file of version V2.00, V2.01 to V2.02.

・There is no initialization of equipment configuration due to version upgrade.
Also, repeated calibration or inspection due to the upgrade are not required.
 

Version Upgrade History

V2.02 (May 2022)
Fixed the following
  - The first measurement value may be out of the accuracy when the range changes to 30m, 300m or 3Ω by the auto range action.

V2.01 (August 2018)
The following was added
  - Increased resistance to damage due to power failure while rewriting settings

V2.00 (October 2015)
Changed MAIN CPU
Fixed the following

  - Communication commands special form of numbers may not be correctly determined
  - Panel load dialog may not be displayed on panel screen
  - "READ?"  may not respond immediately after setting.
  - Error recovery may not work correctly with communication commands
  - External trigger measurement may cause double trigger measurement.
  - When saving and renaming a 10-character panel name after clearing the panel, garbage may appear in the panel name query.
Added the following
  - Added a function to mute all buzzer sounds
  - Added measurement lead bug to the fuse error display.

------

V1.04 (June 2022)
Fixed the following
  - The first measurement value may be out of the accuracy when the range changes to 30m, 300m or 3Ω by the auto range action.

V1.03 (December, 2013)
Fixed the following
  - Power frequency may not be set correctly when power frequency is set by communication command during measurement operation.
  - Trigger application just before the setting is completed may cause malfunctions due to triggering that should be inhibited.
  - Auto ranging is forcibly turned off when the comparator is switched from ON to OFF setting.
  - Even if the EOM signal of EXT I/O is set to pulse, the EOM signal may not pulse but become hold output during zero adjustment.
  - Even if the EOM signal of EXT I/O is set to pulse, the EOM signal is not pulsed but output to hold when measuring with the "READ?" or "INIT" command.
  - If an external trigger is applied at the timing of entering the setting screen, measurement cannot be performed for 3 seconds.
  - LCD backlight does not turn off automatically when external trigger is set by communication command after 1 minute or more in internal trigger state.
  - When the LCD backlight brightness is set to 0, the backlight may become bright when the backlight is automatically turned off.
  - Trigger measurement is not performed when "*TRG" command is sent immediately after range setting by communication command.
  - When communicating in the state of system error at startup, the screen remains error, but the measurement operates.
  - When TRIG signal and KEY_LOCK signal of external I/O are input at the same time, the trigger occurs twice and measurement is performed twice.
  - When a large amount of data is sent via RS232C, data may be dropped.
  - Error position may not be displayed on the communication monitor when a communication error occurs.
  - When a query with seven or more command delimiters is passed as a communication command, the instrument hangs.
  - When auto-range is performed with auto-hold ON, the range change may be mistaken.
The following was added
  - Add judgment delay function.
  - Added data output condition function.
  - Addition of function to automatically output data after auto-hold
  - Addition of a function to automatically output data only when the judgment result is set
  - Added a function to print at any timing during interval printing.
  - Added a function to change the zero adjustment range.

V1.02 (December 2012)
First release
 

Operating Environment

Requires a windows 10 PC.