CHARGE UNIT 8947
Instantly measure acceleration waveform with pick-up sensor
Discontinued
Key Features
- 2 ch, Charge-output type piezoelectric acceleration pick-up sensor, Acceleration pick-up sensor with an internal preamp
- 12-bit A/D resolution measurement
- Buit in Anti-aliasing filter for FFT analysing
- Isorated input, up to 30 V rms AC, 60 V DC (maximum rated voltage to earth)
Model No. (Order Code)
8947 | Discontinued, For the 8860-50, 8861-50, and similar products |
---|
Note: Use with the Memory HiCorder. The charge-output type piezoelectric acceleration pick-up sensor, and acceleration pick-up sensor with an internal preamp is not provided by Hioki, and must be purchased from a separate vendor.
Basic specifications
Accuracy guaranteed: 1 year
Measurement functions | Number of channels: 2, for acceleration measurement | |||||
---|---|---|---|---|---|---|
Input connectors | Voltage input/integrated preamplifier input: metallic BNC connector (for voltage input: input impedance 1MΩ, input capacitance 200pF or less) Charge input: miniature connector (#10-32 UNF) Max. rated voltage to earth: 30Vrms or 60V DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
|||||
Suitable transducer | Charge input: Charge-output type piezoelectric acceleration pick-up sensor Internal preamp input: Acceleration pick-up sensor with an internal preamp |
|||||
Measurement range Charge input (miniature connector) Internal pre-amp input (BNC connector) |
50m (m/s2)/DIV to 10k (m/s2)/DIV, 12 ranges × 6 types, charge input sensitivity: 0.1 to 10 pC/(m/s2), integrated pre-amplifier input: 0.1 to 10 mV/(m/s2), amplitude accuracy: ±2% of full scale, frequency characteristics: 1 to 50kHz, +1/-3dB, low-pass filter: 500Hz/5kHz, pre-amplifier drive power source: 2mA ±20%, +15V ±5%, maximum input charge: ±500pC (high-sensitivity setting, 6 ranges), ±50000pC (low-sensitivity setting, 6 ranges) | |||||
Measurement range Voltage input (BNC connector) |
500μV to 2V/DIV, 12 ranges, DC amplitude accuracy: ±0.4% of full scale, frequency characteristics: DC to 400kHz, +1/-3 dB, low-pass filter: 5Hz/500Hz/5kHz/100kHz, input coupling: DC, GND, AC, Max. allowable input: 30Vrms or 60V DC | |||||
Measurement resolution | 1/80 to 1/32 of measurement range (depending on measurement sensitivity; installed in 8860 series) | |||||
Highest sampling rate | 1MS/s (2-channel simultaneous sampling) | |||||
Anti-aliasing filter | Integrated filter for suppressing aliasing distortion caused by FFT processing (automatic cutoff frequency setting/OFF) | |||||
Dimensions and mass | 170 mm (6.69in)W × 20 mm (0.79in)H × 148.5 mm (5.85in)D, 310 g (10.9 oz) | |||||
Included accessories | None |