How to make my own probe for resistance testers?
QI want to make probes to install a resistance meter into an automation test system. Is there anything to be noted?
A
The cable section is shielded, so the following should be noted when you make your own probes.
Place shielding connected to SOURCE-L.
Cable length should be within 5 meters (using a cable material resistance within 100mΩ/m)