16ch SCANNER UNIT 8958
Measure minute voltage or temperature waveform by 16 bit resolution, isolated input
Discontinued
Key Features
- 16-bit A/D high-resolution measurement
- 16 channels scan input
- Isorated input, up to 33 Vrms AC (maximum rated voltage to earth)
Model No. (Order Code)
8958 | Discontinued, For the 8860-50, 8861-50 |
---|
Note: This module must be used with the Memory HiCorder 8860 series. Thermocouple are not included. Please purchase them separately.
Basic specifications
Accuracy guaranteed: 1 year
Measurement functions | Number of channels: 16, for voltage measurement/temperature measurement with thermocouple |
|||||
---|---|---|---|---|---|---|
Input connectors | Voltage input/Thermocouple input: screw-type terminal strip, recommended wire diameter *1, detachable terminal block (with cover) *1 Recommended cable, single-wire: 0.14 to 1.5 mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16 Input impedance: 1 MΩ, 850 kΩ with line fault detection ON, Max. rated voltage to earth: 33 V rms or 70 V DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
|||||
Voltage measurement range | 5m, 50m, 500m, 2V/DIV, 4 ranges, full scale: 20DIV, measurement range: ±100% of full scale, digital filter: 10Hz/50Hz/60Hz, measurement resolution 1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860 series) | |||||
Temperature measurement range (Upper and lower limit values depend on measurement input range of sensor) |
10°C/DIV (-100°C/ to +200°C), 50°C/DIV (-200°C/ to +1000°C), 100°C/DIV (-200°C/ to +2000°C), 3 ranges, full scale: 20DIV, digital filter: 10Hz/50Hz/60Hz, measurement resolution 1/1000 of measurement range (using 16-bit A/D conversion; installed in 8860 series) | |||||
Thermocouple range (JIS C 1602-1995) (ASTM E-988-96) |
K: -200 to 1350°C, J: -200 to 1200°C, E: -200 to 1000°C, T: -200 to 400°C, N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C, B: 400 to 1800°C, W (WRe5-26): 0 to 2000°C, reference junction compensation: internal/ external (switchable), line fault detection ON/OFF switchable | |||||
Data refresh rate | 50ms/all channels (digital filter OFF), 300ms/all channels (digital filter 50Hz/60Hz), 1.4 s/all channels (digital filter 10Hz) | |||||
Accuracy | Voltage: ±0.2% of full scale, thermocouple (K, J, E, T, N): ±0.05% of full scale ±1°C, (R, S, B, W): ±0.05% of full scale ±2°C (400°C or more), ±0.05% of full scale ±3.5°C (less than 400°C), reference junction compensation accuracy: ±1°C (added to measurement accuracy with internal reference junction compensation) |
|||||
Max. allowable input | 40V DC (the maximum voltage that can be applied across input pins without damage) | |||||
Dimensions and mass | 170 mm (6.69in)W × 20 mm (0.79in)H × 183 mm (7.20in)D, 385 g (13.6 oz) | |||||
Included accessories | Flathead screwdriver × 1, short bar × 2 |