Measuring intermittent leakage current with PQA

QCan I measure intermittent leakage current with the PQ3198, PQ3100, PW3365 or PW3360?

A

To capture intermittent leakage current signals, you need an instrument that can perform gapless continuous measurement.  This specification is featured in the PQ3198 and PQ3100 Power Quality Analyzers, as well as the PW3365 and PW3360 Clamp On Power Loggers.  These are the specifications:

RMS: Continuous measurement approximately every 200ms
Inrush current: Continuous measurement every half wave

RMS: Continuous measurement approximately every 200ms