3281 peak hold function

QThe 3281 and 3282 also have a peak function. Is it possible to capture the crest value for a rush current as with the 3284 and 3285?

A

Although the 3281 and 3282 offer peak hold capability through the combination of peak measurement and recording measurement, these instruments are not able to capture crest values 100% of the time. The 3281 updates the data every 250 ms. Measurement takes 128 ms, but calculation and display processing take up the remaining time, raising the possibility that a peak might be missed.