INSULATION / WITHSTANDING HiTESTER 3159

0 to 0.2 kV/5 kV AC Insulation and Hipot Tester

Key Features

  • Continuous testing of insulation (500 V/1000 V) and withstand voltage (500 VA transformer capacity)
  • Insulation to withstand series test or withstand to insulation series test in auto mode, or individual test in manual mode
  • Save up to 10 test settings each for the withstanding and insulation testing modes
  • External I/O, RS-232C interface, Status output (relay contacts)

Model No. (Order Code)

3159 For 100 V AC power supplies
3159-02 For 220 V AC power supplies only

Basic Specifications

Accuracy guaranteed: 1 year
Monitor function Output voltage, detected current, insulation resistance, Refresh rate: 2 times/s
Power supply 100 V AC, (50/60 Hz), 800 VA max.
Dimensions and mass 320 mm (12.60 in) W × 155 mm (6.10 in) H × 330 mm (12.99 in) D,
21.5 kg (758.4 oz)
Included accessories H.V. Test Lead 9615 (high voltage side and return, 1 each) ×1, Power cord ×1, Instruction manual ×1, Spare fuse ×1

[Withstanding test]

Testing voltage 0 to 2.5 kV / 0 to 5.0 kV AC, 2 range configuration 500 VA (30 minutes rated)
Voltage setting Manual setting
Waveform/Frequency Same as the power supply waveform, synchronized with the power supply
Current measurement 0.01 mA to 120 mA, True RMS rectified (digital display)
Voltage meter Accuracy: ±1.5 % f.s. (digital), ±5 % f.s. (analog, f.s.=5 kV)
Current measurement 0.01 mA to 120 mA, (Average value rectified, effective value digital display)
Measurement range 2 mA/8 mA (0.01 mA resolution), 32 mA (0.1 mA resolution), 120 mA (1 mA resolution)
Voltage meter Digital, Accuracy: ±1.5 % rdg. (f.s.=5.00 kV)
Judgment function Window comparator method (Digital setting)

[Insulation test]

Testing voltage 500 V DC, 1000 V DC, Unloaded voltage: 1 to 1.2 times rated voltage
Rated testing current 1 to 1.2 mA, Shorted current: 4 to 5 mA (at 500 V), 2 to 3 mA (at 1000 V)
Measurement range/accuracy 0.5 MΩ to 999 MΩ (at 500 V), and 1 MΩ to 999 MΩ (at 1000 V): ±4 % rdg.,
1000 MΩ to 2000 MΩ: ±8 % rdg.
Judgment function Window comparator method (Digital setting)

[Timer]

Setting range 0.5 to 999 s
-- --

Input/Output cords (4)