10:1PROBE 9665
Measures high voltages by attenuating the voltage to a lower level
The 9665 is 10:1 attenuation probes used to measure high voltage with Hioki Memory HiCorders.
Key Features
- Attenuates the input voltage to 1/10
- Input capacitance to the probe can be lowered to reduce the load effect
Model No. (Order Code)
9665 | For the Memory HiCorder series |
---|
Note: This probe does not expand the maximum rated voltage above ground of an isolated input.
*Please check for Memory HiCorder compatibility by referring to the recorder's product catalog.
Basic specifications
Division ratio | 10 : 1 ±1% (at 100 V DC) (use with a device having a 1 MΩ input resistance) |
|||||
---|---|---|---|---|---|---|
Maximum input voltage | 1 kV rms (500 kHz or less) | |||||
Max. rated voltage to earth | 1000 V CAT II * * Maximum rated voltage above ground depends on input unit. This probe does not expand the maximum rated voltage above ground of an isolated input. |
|||||
Input Impedance | 10 MΩ | |||||
Frequency characteristics | DC to 150 MHz | |||||
Input capacitance | 18 pF ±2 pF | |||||
Compensation capacitance range (Equipment side) |
15 to 45 pF | |||||
Cable length | 1.5 m (4.92 ft) | |||||
Included accessories | Vinyl case ×1 |