DIGITAL MULTIMETER DT4282

World's Premier Digital Multimeter! Superior Accuracy and High Response, Topped with Safety Terminal Shutters

The Hioki DT4200 Digital Multimeter Series consist of a complete lineup of 9 models ranging from professional to industrial to pocket types. The DT4282 is a high-precision 60,000 count digital multimeter with direct 10A input and all of the functions and features required for advanced testing in laboratories and R&D.

CAT IV 600 V, CAT III 1000 V

Video on how to use digital multimeters

This video introduces how to use the DT4280 series of digital testers as well as some of the instrument’s featured functionality. Hioki testers are engineered to help professional electricians do their jobs safely and quickly.

Key Features

  • 60000 count, 5-digit display, high-resolution measurements
  • ±0.025% DC V basic accuracy, wide 20 Hz to 100 kHz AC V frequency characteristics
  • Low-pass filter cuts high harmonics (when measuring inverter fundamental waveforms)
  • Terminal shutter mechanism (prevents erroneous test lead insertion)
  • Measure up to 10 A with direct input (DT4282 only)
  • Dual display lets you check voltage and frequency simultaneously

Model No. (Order Code)

DT4282 10 A direct input

The ultimate digital multimeter, ideal for use R&D applications

The DT4282, which is part of Hioki’s DT4280 series of high-end digital multimeters, is designed for use in laboratories and R&D applications where you wish to measure a wide variety of parameters.

- DC current 600.00 μA / 6000.0 μA / 60.000 mA / 600.00 mA /6.0000 A / 10.000 A
- AC current 600.00 μA / 6000.0 μA / 60.000 mA / 600.00 mA /6.0000 A / 10.000A
- Conductance 600.00 nS

Measure output voltage on the secondary sides of inverters by eliminating harmonic components

Accurately measure the fundamental wave alone by eliminating harmonic components with the DMM’s low-pass filter function.

Measure ripple voltage of DC charging systems

You can measure ripple voltage with a superposed DC signal by using peak value measurement and DC + AC voltage measurement.

Percentage display for 4 to 20 mA instrumentation signals

The digital multimeters can convert and display instrumentation signals of 4 to 20 mA or 0 to 20 mA representing properties such as temperature, pressure, and flow rate from a transducer or other device that converts sensor data into an electrical signal.

600.00 μA range for measuring very low currents used by gas-burning devices

Accurately measure the very low currents such as burner flame currents.

True RMS measurement for accurate measurement of even distorted current waveforms

Digital multimeters implement two types of measurement: mean-value measurement and true RMS measurement. The two methods yield significantly different measurement results for current waveforms, which are often distorted. For accurate measurement, it is necessary to use an instrument with true RMS capability.

Measurement data can be downloaded to, reviewed on, and saved to a computer.

If you connect the DT4282 to a computer with the optional Communication Package DT4900-01, you can graph data stored in the instrument’s internal memory and display measurement results in real time on the computer. In addition, displayed measured values can be saved (in text format).

*Use of optical communications keeps the computer and instrument electrically isolated to ensure safety.
*The computer and multimeter are electrically isolated by means of optical communications so that data can be sent without concern for electrical hazards.

Designed and manufactured in Japan to ensure high quality and guaranteed with a 3-year warranty for peace of mind

All development, design, and manufacturing processes for almost all Hioki digital multimeters are carried out at our Head Office in Nagano Prefecture. Some of the industry’s most advanced technological capabilities enable us to deliver products of the highest possible quality.

Safe measurement requires use of an instrument that suits the measurement location.

To ensure operators’ ability to use measuring instruments safely, IEC 61010 classifies the locations in which instruments are used into a series of safety-based measurement categories (ranging from CAT II to CAT IV). Using an instrument that does not satisfy the required safety level can lead to an electrical accident.

Basic specifications

Accuracy guaranteed: 1 year
Regarding DMM Accuracy: Due to the many ranges and functions available in a DMM, only the basic accuracy is indicated for reference. Please refer to the brochure for detailed accuracy information.
DC Voltage range 60.000 mV to 1000.0 V, 6 ranges, Basic accuracy: ±0.025 % rdg. ±2 dgt.
AC Voltage* range 60.000 mV to 1000.0 V, 6 ranges, Frequency characteristics: 20 Hz - 100 kHz
Basic accuracy 45 - 65 Hz : ±0.2 % rdg. ±25 dgt. (True RMS, crest factor 3)
DC + AC Voltage* range 6.0000 V to 1000.0 V, 4 ranges, Frequency characteristics: 20 Hz - 100 kHz
Basic accuracy 45 - 65 Hz : ±0.3 % rdg. ±30 dgt. (True RMS, crest factor 3)
Resistance range 60.000 Ω to 600.0 MΩ, 8 ranges, (Conductance: 600.00 nS, DT4282 only)
Basic accuracy: ±0.03 % rdg. ±2 dgt.
DC Current range 600.00 μA to 10.000 A, 6 ranges, Basic accuracy: ±0.05 % rdg. ±5 dgt.
AC Current* range 600.00 μA to 10.000 A, 6 ranges,
Basic accuracy 45 - 65 Hz : ±0.6 % rdg. ±5 dgt. (True RMS, crest factor 3)
Frequency characteristics: 20 Hz - 20 kHz (at 600 μA to 600 mA range)
AC Current* range
(use with Clamp on probes)
N/A
Peak DC V measurement: Signal width 4 msec or more (single), 1 msec or more (repeated)
AC V, DC/AC A measurement:
Signal width 1 msec or more (single), 250 μsec or more (repeated)
Capacitance range 1.000 nF to 100.0 mF, 9 ranges, Basic accuracy: ±1.0 % rdg. ±5 dgt.
Continuity Check Continuity threshold: 20/50/100/500 Ω, Response time: 10 ms or more
Diode test Open terminal voltage: 4.5 V or less, Testing current 1.2 mA or less, Threshold of forward voltage: 0.15 V to 3 V, seven stages
Frequency range AC V, DC+AC V, AC A measurement, at pulse width 1 μs or more (50 % duty ratio)
99.999 Hz (0.5 Hz or more) to 500.00 kHz, 5 ranges, ±0.005 % rdg. ±3 dgt.
dB conversion Standard impedance setting (dBm), 4 Ω to 1200 Ω, 20 stages
Display dB conversion value of AC voltage (dBV)
Temperature
(thermocouples)
K: -40.0 °C to 800.0 °C (-40.0 °F to 1472.0 °F)
Add accuracy of the Thermocouple probe to main unit accuracy: ±0.5 % rdg. ±3 °C
Other functions Filter function (remove harmonic noise, use only at 600 V AC, 1000 V AC ranges), display value hold, auto hold, MAX/MIN value display, PEAK value display, relative value display, sampling setting, internal memory (400 data), auto-power save, USB communication (option), mis-insertion prevention shutters, decibel conversion, 4-20 mA percentage conversion
Display Main and Sub displays: 5-digits LCD, max. 60000 digits
Display refresh rates 5 times/s (Capacitance measurement: 0.05 to 2 times/s, depending on measured value, Temperature: 1 time/s )
Power supply LR6 (AA) alkaline batteries ×4, Continuous use: 100 hours
Dimensions and mass 93 mm (3.66 in) W × 197 mm (7.76 in) H× 53 mm (2.09 in) D, 650 g (22.9 oz) (with test leads holder and batteries)
Included accessories Test Lead L9207-10 ×1, Instruction Manual ×1, LR6 Alkaline Battery ×4
  • *Zero-suppression: For small inputs below the guarantee range, zero is effectively displayed.

Test leads (3)

L9207-10 is a bundled accessory

Options for Test leads (for the L9207-10, L9300) (2)

Attach L9207-10 with the cap removed.
Slide the guard of L9300 and attach as measurement category II.

Options for Test leads (for the L4930) (9)

Temperature measurement (1)

Other options (4)


Related Products