STRAIN UNIT U8554
Strain measurement with 1 ms sampling
Key Features
- 5 channels, strain measurement
- Sampling rate of up to 1 ms with 5 channels
- Direct input of strain gages and strain gage transducers
Model No. (Order Code)
U8554 | For the LR8450, LR8450-01 |
---|
U8554 cannot perform measurement by itself. Memory HiLogger LR8450 or LR8450-01 is required separately.
Basic specifications
Accuracy guaranteed: 1 year
Connectable instruments | Memory HiLogger LR8450 (Designed for use with plug-in units only) Memory HiLogger LR8450-01 (Wireless LAN model) |
|||||
---|---|---|---|---|---|---|
Operating temperature and humidity |
-10°C to 50°C, 80% RH or less (no condensation) | |||||
Storage temperature and humidity |
-20°C to 60°C, 80% RH or less (no condensation) | |||||
Vibration resistance | JIS D 1601:1995 5.3(1), Class 1A (passenger vehicle) equivalent | |||||
Included accessories | Instruction manual × 1, mounting screw × 2, wiring confirmation label | |||||
Number of input channels |
5 ch (Able to set voltage or strain for each channel.) | |||||
Input terminals | Push-button type terminal block (5 terminals per channel), outfitted with terminal block cover Set DIP switches according to measurement target. |
|||||
Measurement target |
Voltage Strain: Strain gage 1-gage method (2-wire setup), 1-gage method (3-wire setup), 2-gage method (adjacent sides), 4-gage method Strain gauge transducer |
|||||
Input type | Balanced differential input, Simultaneous sampling of all channels (non-isolated channels) |
|||||
Voltage measurement (resolution) | 1 mV f.s. (50 nV), 2 mV f.s. (100 nV), 5 mV f.s. (250 nV), 10 mV f.s. (500 nV), 20 mV f.s. (1μV), 50 mV f.s. (2.5μV), 100 mV f.s. (5μV), 200 mV f.s. (10μV) | |||||
Strain measurement (resolution) | 1,000 με f.s. (0.05 με), 2,000 με f.s. (0.1 με), 5,000 με f.s. (0.25 με), 10,000 με f.s. (0.5 με), 20,000 με f.s. (1 με), 50,000 με f.s. (2.5 με), 100,000 με f.s. (5 με), 200,000 με f.s. (10 με) |