RGB LASER METER TM6102

Definitively Measure the White in Laser Displays - Specially Designed for RGB Lasers

Discontinued

The RGB LASER METER TM6102 evaluates the white balance of displays that use lasers by measuring centroid wavelength and optical power, which are key characteristics in the evaluation of laser light sources, using the proprietary Discrete Centroid Wavelength Method, a new technique for directly measuring the centroid wavelength and optical power of laser light sources. Applications include processes designed to adjust and inspect RGB lasers used in automotive head-up displays and head-mounted displays by measuring radiant illumination, centroid wavelength, illuminance, and chromaticity.

Achieve Both the Measurement Performance of a Spectrometer and the Convenient Usability of a Colorimeter in a Single Device

The Hioki TM6102, TM6103 and TM6104 RGB laser light measurement instruments measures both the centroid wavelength and optical power of each wavelength of a RGB composite at the same time. Set the tolerance levels for photometric quantity and chromaticity to the target range of to automatically calculate the adjustment range

Set the target and tolerance range of photometric quantity and chromaticity levels and and the revolutionary devices will use the innovative White Balance Adjustment Assistance function to match the specified range of each RGB wave’s level of emission.

Key Features

  • Proprietary Discrete Centroid Wavelength Method for laser photometry
  • Cut adjustment time in half with white balance navigation
  • Measure up to 9 points on 12-inch* screen simultaneously
  • Modulated light function for displays with a wide color gamut
  • Low incidence angle dependence in chromaticity
  • *12 inches: 4:3 250 mm × 180 mm (9.84 in x 7.09 in)

Model No. (Order Code)

TM6102 Illuminance

A comment from the Good Design Award 2017

“Hioki invented the Discrete Centroid Wavelength Method for instruments used to measure laser light sources, a technology that promises to boost productivity in the manufacture of displays by speeding up the white balance adjustment process, which has previously required skill and experience. This award recognizes a waste-free, functional design that allows the instrument to be easily aligned with the laser’s optical axis from the side simply by aligning a mark on the enclosure with the beam.”

Basic specifications

Accuracy guaranteed: 1 year
Measurement object Laser light
Note: Accuracy is guaranteed with a laser light source. Accuracy is not guaranteed with standard illuminant A (light bulb).
Measurement parameters Irradiance, illuminance, centroid wavelength

Tristimulus values XYZ, chromaticity (xy, u’v’), correlated color temperature, delta uv, dominant wavelength, NTSC ratio, white balance target value of radiometric quantity
Radiometric quantity Irradiance
Measurement range 0.0002 to 200 [W/m^2]
Relative accuracy ±4.6% rdg. (473 nm, 40 μW)
Standard (532 nm, 60 μW)
±4.6% rdg. (633 nm, 80 μW)
Accuracy ±6.5% rdg. (532 nm, 9 mW/m^2)
Photometric quantity Illuminance
Measurement range 0.2 to 110 000 [lx]
Centroid wavelength measurement range Blue : 435 nm to 477 nm, Green : 505 nm to 550 nm, Red : 615 nm to 665 nm
White balance adjustment assistance functions (Set parameters) Target value of photometric quantity, tolerance of photometric quantity, target value of chromaticity (x, y), tolerance of chromaticity (x, y)
Interfaces LAN (TCP/IP) * A display is not available on the unit.
Power supply AC ADAPTER: Z1008 (100 V AC to 240 V AC, 9.5 VA)
Dimensions and mass 65 mm (2.56 in) W × 83 mm (3.27 in) H × 126 mm (4.96 in) D, 700 g (24.7 oz)
Included accessories AC ADAPTER: Z1008 ×1, Power cord ×1, Light shielding cap ×1, LAN cable (3 m, 9.84 ft length), Instruction manual ×1, Application disk CD-R (RGB Laser Utility application program) ×1