DIGITAL VOLTMETER UNIT U8991
For minute voltage measurement, 6 1/2-digit high resolution: Max 1200000 counts
Key Features
- DVM unit dedicated to 4-channel DCV measurement, 108 channels simultaneous sampling in combination with MR8740T
- High-precision measurement of fine voltage fluctuation investigation etc. of sensor output
- High accuracy of ±0.02 % and super-high 1.2 million-count resolution
- High-speed sampling at 50 samples/s
- Isolated input (channel-to-channel and channel-to-chassis: max. rated voltag to earth of 100 V AC/DC)
Model No. (Order Code)
U8991 | For the MR8740-50 |
---|
List of Available Memory HiCorders
Please refer to the table here for the types of input units that can be used for each Memory HiCorder.
Basic specifications
Measurement functions | Number of channels: 4, for DC voltage measurement | |||||
---|---|---|---|---|---|---|
Input terminals | Isolated BNC connectors (Input impedance: 100 MΩ or higher with 1 V f.s. to 10 V f.s. range, otherwise 10 MΩ) Max. rated voltage to ground: 100 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
|||||
Measurement range | 1, 10, 100 V f.s., 3 ranges | |||||
Measurement resolution | 1/1 000 000 of measurement range (using 24-bit ΔΣ modulation A/D) | |||||
Integration time | 20 ms × NPLC (during 50 Hz), 16.67 ms × NPLC (during 60 Hz) | |||||
Basic measurement accuracy | ±0.02% rdg. ±0.0025% f.s. | |||||
Maximum input voltage | 100 V DC (maximum voltage that can be applied across input pins without damage) | |||||
Dimensions and mass | 106 (4.17 in) W ×19.8 (0.78)H × 196.5 (7.74 in)D mm, 250 g (8.8 oz) | |||||
Included accessories | None |
Input cable (A) (4)
*Voltage is limited to the specifications of the iput section
• Flexible φ 4.1 mm (0.16 in) thin dia.
• Cable allowing for up to 600 V input
• 1.8 m (5.91 ft) length
Red/black set attaches to the ends of the Connection Cord L9790
Red/black set attaches to the ends of the Connection Cord L9790
Red/black set attaches to the ends of the Connection Cord L9790
Input cable (B) (1)
*Voltage is limited to the specifications of the iput section
•φ 5.0 mm (0.20 in) dia.
•Cable allows for up to 300 V input
•1.7 m (5.58 ft) length
•Small alligator clip