LOGIC PROBE 9320-01
Detect the Presence of Voltage or Relay Contact Signals with Memory HiCorders
Key Features
- 4 input channels (Not isolated)
- For ON/OFF detection of voltage/contact signals
- Select from 3 steps of input thresholds
- Response pulse width of 500 ns or more
- Compatible with Memory HiCorders featuring a miniature terminal
Model No. (Order Code)
9320-01 | For the Memory HiCorder, miniature terminal type |
---|
Note: Verify that your Memory HiCorder features a miniature connection terminal before purchase
Basic specifications
Note: The unit-side plug of the 9320-01 is different from the 9320
Function | Detection of voltage signal or relay contact signal for High/Low state recording | |||||
---|---|---|---|---|---|---|
Input | 4 channels (common ground between unit and channels), digital/contact input, switchable (contact input can detect open-collector signals) Input resistance: 1 MΩ (with digital input, 0 to +5 V) 500 kΩ or more (with digital input, +5 to +50V) Pull-up resistance: 2 kΩ (contact input: internally pulled up to +5 V) |
|||||
Digital input threshold | 1.4V/ 2.5V/ 4.0V | |||||
Contact input detection resistance | 1.4 V: 1.5 kΩ or higher (open) and 500 Ω or lower (short) 2.5 V: 3.5 kΩ or higher (open) and 1.5 kΩ or lower (short) 4.0 V: 25 kΩ or higher (open) and 8 kΩ or lower (short) |
|||||
Detectable pulse width | 500 ns or longer | |||||
Max. allowable input | 0 to +50V DC (the maximum voltage that can be applied across input pins without damage) | |||||
Dimensions and mass | 62 mm (2.44 in)W × 94 mm (3.70 in)H × 20 mm (0.79 in)D, 150 g (5.3 oz), Main unit cable 1.5 m (4.92 ft), input section cable 300 mm (0.98 ft) | |||||
Included accessories | IC clip leads ×1, Alligator clip leads ×1, Carrying case ×1, Instruction Manual ×1 |