HIGH SPEED ANALOG UNIT U8976

High speed 200 MS/s, isolated input

Key Features

  • High-speed sampling of 200 MS/s ensures catching high-speed phenomena
  • Isolated 2 channel input
  • Max. rated voltage to ground: 1000 V AC or DC (CAT II)
  • High resolution measurement with 12 bit A/D
  • For Memory HiCorder MR6000 only

Model No. (Order Code)

U8976 For the MR6000 and similar products
Note: Use it by installing it in the Memory HiCorder. Input code kind is not included so purchase separately.

Blazingly fast sampling that won't fail

The High Speed Analog Unit U8976 delivers a 30 MHz frequency band in addition to high-speed sampling at 200 MS/s. It has the performance needed to accurately capture switching waveforms during inverter testing, an application where high efficiency is critical. Adapted to the Memory HiCorder's direct input feature, it can accept inputs of up to 400 V DC.

Accurately measure inrush current with high-speed sampling

Combine the High-Speed Analog Unit U8976’s 30 MHz frequency band with the Current Probe CT6711 to measure inrush currents and minuscule currents.
Power can be supplied to current probes by using the Power Probe Unit Z5021 (MR6000 only).

List of Available Memory HiCorders

Please refer to the table here for the types of input units that can be used for each Memory HiCorder.

List of Available Memory HiCorders

Basic specifications

Accuracy guaranteed: 1 year
Measurement functions Number of channels: 2, for voltage measurement
Input terminals Isolated BNC terminal (Input impedance: 1 MΩ, Input capacitance: 22 pF)
Max. rated voltage to earth: 1000 V AC or DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage)
Measurement range 100, 200, 400 mV f.s., 1, 2, 4, 10, 20, 40, 100, 200, 400 V f.s., 12 ranges
(AC voltage for possible measurement/display using the memory function: 280 V rms)
Low-pass filter: 5/500/5 k/1 MHz
Measurement resolution 1/1600 of measurement range (using 12-bit A/D conversion)
Highest sampling rate 200 MS/s (2 channel simultaneous sampling)
Measurement accuracy ±0.5% f.s. (with filter 5 Hz, zero position accuracy included)
Frequency characteristics DC to 30 MHz -3 dB, with AC coupling: 7 Hz to 30 MHz -3dB
Input coupling AC/DC/GND
Max. allowable input 400 V DC (when direct input), 1000 V DC (when using the 9665)
Dimensions and mass 106 mm (4.17 in) W × 19.8 mm (0.78 in) H × 196.5 mm (7.74 in) D, 280 g (9.9 oz)
Included accessories None

Use with these Products (1)

* Main products are listed below. For more information, please see each product page.

Voltage measurement Type A (4)

* Input voltage is limited to the specifications of the input modules in use

* Max. rated voltage to earth is limited to the specifications of the input modules in use

Voltage measurement Type B (1)

* Input voltage is limited to the specifications of the input modules in use

* Max. rated voltage to earth is limited to the specifications of the input modules in use

Voltage measurement Type C (2)

* Input voltage is limited to the specifications of the input modules in use

* Max. rated voltage to earth is limited to the specifications of the input modules in use

Voltage measurement Type D (2)

* Max. rated voltage to earth is limited to the specifications of the input modules in use

* For a list of compatible Memory HiCorders, please see the product catalog