DC SIGNAL SOURCE SS7012

DC Voltage and Current Signal Generator and Tester for Routine Maintenance of Standard Devices

The Hioki SS7012 is a compact DC signal source and calibrator that is used to perform maintenance or loop tests of instrumentation systems and can be operated from a computer via a USB cable. It can also be used to calibrate thermometers and measure DC voltage and DC current.

Key Features

  • Improve stability and reduce calibration costs compared with the previous HIOKI model
  • For instrumentation systems (4 - 20 mA) and loop testing
  • Check temperature control equipment and electric distribution
  • 8 types of thermocouples to test thermoelectric power generation
  • Ideal for electrical device evaluating and routine maintenance of production equipment such as calibrators
  • Use the max. 25 mA DC sink as an electric load

Model No. (Order Code)

SS7012
Use of the AC Adapter and /or rechargeable batteries and dedicated charger is recommended.

Basic specifications [Generation functions]

Accuracy guaranteed: 1 year
Circuit method Bipolar sink and source
Constant voltage 2.5 V: 0 to ±2.5000 V (±0.03 % of setting ±300 μV, 100 μV resolution)
25 V: 0 to ±25.000 V (±0.03 % of setting ±3 mV, 1 mV resolution)
Constant current 25 mA: 0 to ±25.000 mA (±0.03 % of setting ±3 μA, 1 μA resolution)
Thermoelectric power generation K: at TC: 0 ˚C, -174.0 to 1372.0 ˚C (±0.05 % of setting ±0.5 ˚C, 0.1 ˚C resolution), Other types: E, J, T, R, S, B, N selectable
Thermoelectric power generation K: at TC: RJ, -174.0 to 1372.0 ˚C (±0.05 % of setting ±1.0 ˚C, 0.1 ˚C resolution), Other types: E, J, T, R, S, B, N selectable
Standard resistance (Rs) 100 Ω (±0.2 Ω)
Automatic generation Number of memory steps: 20, Interval time: 1 to 99 sec (at CV, CC, TC mode)

[Measurement functions]

Voltage 2.5 V: 0 to ±2.8000 V (±0.03 % rdg. ±300 μV, 100 μV resolution, 1 MΩ input resistance)
25 V: 0 to ±28.000 V (±0.03 % rdg. ±3 mV, 1 mV resolution, 1 MΩ input resistance)
Current 25 mA: 0 to ±28.000 mA (±0.03 % rdg. ±3 μA, 1 μA resolution, smaller than 30 Ω input resistance)
Temperature -25.0 to 80.0 ˚C (±0.5 ˚C at 23 ±5 ˚C, 0.1 ˚C resolution, use with the RJ sensor 9184)
Sampling rate Approx. 1.67 times/sec

[General functions]

Additional functions Zero adjustment, Overflow display, USB communication, Monitor
Power supply AC adapter 9445-02/-03 (100 to 240 V AC 50/60 Hz, 9 VA),
Ni-MH battery HR6 × 4, 6 VA, (fully charged 2500 mAh Ni-MH batteries: 170 minutes continuous use), or LR6 (AA) alkaline battery × 4, 6 VA
Dimensions and mass 104 mm (4.09 in) W × 180 mm (7.09 in) H × 58 mm (2.28 in) D, 660 g (23.3 oz) (including LR6 × 4 batteries)
Included accessories Input cord 9168 ×1, Test lead L9170-10 ×1, Fuse ×1, LR6 (AA) alkaline battery ×4, Instruction manual ×1

Cables (2)

*Included accessories

Options (5)

Commercially available rechargeable batteries (AA Ni-MH batteries ×4) may also be used to power the SS7012. Using locally purchased rechargeable batteries and dedicated battery chargers is recommended; however, HIOKI will not be able to guarantee operatin