VOLTAGE/TEMP UNIT U8550
Voltage, Temperature, and Humidity Measurement with 10 ms Sampling
Key Features
- 15 ch, voltage/temperature (thermocouples)/humidity measurement
- Sampling rate of up to 10 ms with 15 channels
- For DC voltage measurement and temperature measurement using thermocouples
Model No. (Order Code)
U8550 | For the LR8450, LR8450-01 |
---|
U8550 cannot perform measurement by itself. Memory HiLogger LR8450 or LR8450-01 is required separately.
Basic specifications
Accuracy guaranteed: 1 year
Connectable instruments | Memory HiLogger LR8450 (Designed for use with plug-in units only) Memory HiLogger LR8450-01 (Wireless LAN model) |
|||||
---|---|---|---|---|---|---|
Operating temperature and humidity |
-10°C to 50°C, 80% RH or less (no condensation) | |||||
Storage temperature and humidity |
-20°C to 60°C, 80% RH or less (no condensation) | |||||
Vibration resistance | JIS D 1601:1995 5.3(1), Class 1A (passenger vehicle) equivalent | |||||
Included accessories | Instruction manual × 1, mounting screw × 2 | |||||
Number of input channels |
15 ch (Able to set voltage, temperature using the thermocouple, and humidity for each channel) | |||||
Input terminals | M3 screw-type terminal block (2 terminals per channel) | |||||
Measurement target |
Voltage, temperature using the thermocouple, and humidity | |||||
Input type | Scanning by semiconductor relays All channels isolated (Not isolated when measuring humidity) |
|||||
Voltage measurement (resolution) | 10 mV f.s. (500 nV), 20 mV f.s. (1μV), 100 mV f.s. (5μV), 200 mV f.s. (10μV), 1 V f.s. (50μV), 2 V f.s. (100μV), 10 V f.s. (500μV), 20 V f.s. (1 mV), 100 V f.s. (5 mV), 1-5 V f.s. (500μV) | |||||
Temperature measurement (thermocouple) | K: -100.00℃ to 1350.0℃ J: -100.00℃ to 1200.0℃ E: -100.00℃ to 1000.0℃ T: -100.00℃ to 400.0℃ N: -100.00℃ to 1300.0℃ R: 0.00℃ to 1700.0℃ |