TEMP UNIT 8967
Measures temperature waveforms from -200°C to 2000°C
Key Features
- High 16-bit A/D resolution
- Isolated 2 channel input
- Max. rated voltage to ground: 300 V AC, DC
Model No. (Order Code)
8967 | For MR6000, MR8847A, MR8827, and similar products |
---|
Note: This module must be used with the Memory HiCorder. Input cords are not included. Please purchase them separately.
List of Available Memory HiCorders
Please refer to the table here for the types of input units that can be used for each Memory HiCorder.
Basic specifications
Accuracy guaranteed: 1 year
Measurement functions | Number of channels: 2, for temperature measurement with thermocouple (voltage measurement not available) | |||||
---|---|---|---|---|---|---|
Input connectors | Thermocouple input: plug-in connector, Recommended wire diameter: single-wire, 0.14 to 1.5 mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16 Input impedance: min. 5 MΩ (with line fault detection ON/OFF), Max. rated voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
|||||
Temperature measurement range Note: Upper and lower limit values depend on the thermocouple |
10 °C/div (-100 °C to 200 °C), 50 °C/div (-200 °C to 1000 °C), 100 °C/div (-200 °C to 2000 °C), 3 ranges, full scale: 20 div, Measurement resolution: 1/1000 of measurement range (using 16-bit A/D conversion and when installed in the 8847) |
|||||
Thermocouple range (JIS C 1602-1995) (ASTM E-988-96) |
K: -200 to 1350 °C, J: -200 to 1100 °C, E: -200 to 800 °C, T: -200 to 400 °C, N: -200 to 1300 °C, R: 0 to 1700 °C, S: 0 to 1700 °C, B: 400 to 1800 °C, W (WRe5-26): 0 to 2000 °C, Reference junction compensation: internal/ external (switchable), Line fault detection ON/OFF possible |
|||||
Data refresh rate | 3 methods, Fast: 1.2 ms (digital filter OFF), Normal: 100 ms (digital filter 50/60 Hz), Slow: 500 ms (digital filter 10Hz) | |||||
Measurement accuracy | Thermocouple K, J, E, T, N: ±0.1 % of full scale ±1 °C (±0.1 % of full scale ±2 °C at -200 °C to 0 °C), Thermocouple R, S, W: ±0.1 % of full scale ±3.5 °C (at 0 °C to 400 °C or less), ±0.1 % of full scale ±3 °C (at 400 °C or more) Thermocouple B: ±0.1 % of full scale ±3 °C (at 400 °C or more), Reference junction compensation accuracy: ±1.5 °C (added to measurement accuracy with internal reference junction compensation) |
|||||
Dimensions and mass | 106 mm (4.17 in) W × 19.8 mm (0.78 in) H × 204.5 mm (8.05 in) D, 240 g (8.5 oz) | |||||
Included accessories | Ferrite clamp × 2 |