VOLTAGE/TEMP UNIT 8937
Measure voltage or temperature waveform by 12 bit resolution, isolated input
Discontinued
Key Features
- High resolution measurement at voltage input
- Isorated input, up to 30 V AC, 60 V DC (maximum rated voltage to earth)
Model No. (Order Code)
8937 | Discontinued, For the 8860-50, 8861-50, and similar products |
---|
Note: This module must be used with the Memory HiCorder. Thermocouples are not included. Please purchase them separately.
Note: When using Model 8937 with serial number earlier than 041135257 on Models 8861-50/8860-50/8861/8860, residual noise will be 150 μVp-p.
Basic specifications
Accuracy guaranteed: 1 year
Measurement functions | 170 mm (6.69in)W × 20 mm (0.79in)H × 148.5 mm (5.85in)D, 300 g (10.6 oz) | |||||
---|---|---|---|---|---|---|
Input connectors | Voltage input: metallic BNC connector (input impedance 1MΩ, input capacitance 50pF), thermocouple input: plug-in connector (input impedance min. 5.1MΩ), Max. rated voltage to earth: 30Vrms or 60V DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) | |||||
Voltage measurement range | 500μV to 2 V/DIV, 12 ranges, full scale: 20DIV, low-pass filter: 5Hz/500Hz/5kHz/100kHz, Measurement resolution: 1/80 of measurement range (using 12-bit A/D conversion; installed in 8860 series) | |||||
Temperature measurement range | 10°C to 100°C/DIV, 4 ranges, full scale: 20DIV, low-pass filter: 5Hz/500Hz, Measurement resolution:1/80 of measurement range (using 12-bit A/D conversion; installed in 8860 series) | |||||
Thermocouple range | K: -200 to 1350°C, E: -200 to 800°C, J: -200 to 1100°C, T: -200 to 400°C, N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C, B: 300 to 1800°C, Reference junction compensation: internal/ external (switchable) |
|||||
Highest sampling rate | Voltage input: 1MS/s, Temperature measurement: 4kS/s (2-channel simultaneous sampling) | |||||
Accuracy | Voltage input: DC amplitude ±0.4% of full scale, zero position ±0.15% of full scale, Temperature measurement (K, E, J, T, N): ±0.1% of full scale ±1°C, ±0.1% of full scale ±2°C (-200 to 0°C), (R, S): ±0.1% of full scale ±3°C, (B): ±0.1% of full scale ±4°C (400 to 1800°C) , Reference junction compensation accuracy: ±0.1% of full scale ±1.5 °C (internal reference junction compensation) | |||||
Frequency characteristics | Voltage input: DC to 400 kHz +1/-3dB Temperature measurement: DC to 1kHz +1/-3dB |
|||||
Input coupling | DC, GND, AC | |||||
Max. allowable input | 30Vrms or 60V DC (the maximum voltage that can be applied across input pins without damage) | |||||
Dimensions and mass | 170 mm (6.69in)W × 20 mm (0.79in)H × 148.5 mm (5.85in)D, 300 g (10.6 oz) | |||||
Included accessories | None |