How to read "STATUS" data (3169)

How to read "STATUS" data (3169)

Question

How do I read the Status data in the 3169's CSV data?

Answer

The Status is the binary number of the 9 bit data to specify if there is any overrange, overload input (crest factor over) or power outage during time series recording.

Bit       Bit9    Bit8     Bit7     Bit6      Bit5      Bit4     Bit3     Bit2      Bit1       Bit0
Item  PLL     pd        or       ovI4     ovI3     ovI2     ovi1    ovU3     ovU2    ovU1

The Status data is decimal number, so it has to be converted to a 16bit binary number using a scientific calculator