Leakage current measurement using the 3169

Leakage current measurement using the 3169

Question

Is it possible to measure the leakage current using the Clamp on Power HITESTER 3169?

Answer

There is no clamp on sensor selection for the Leak Current Sensors 9657-10 or 9675 on the 3169.
However, you can use them by making the following settings.
    Sensor:  9661 (100mV AC/A)
    CT ratio:  0.01
By observing the fundamental frequency (1st order) captured during harmonic measurement, you can identify leakage current at 50/60Hz.