| Options for Super megohm meter (for Surface resistance or volume resistance measurement) |
|
| SURFACE/VOLUME RESISTANCE MEASUREMENT ELECTRODE SM9001 |
 |

PDF:1.49MB |
- Electrodes compliant with the JIS C 2170 and IEC 61340-2-3 standards
- Measurement voltage up to 1,000 V, and measurement resistance up to 1013 Ω
- Surface and volume resistance of sheets and films can be measured just as they are without the need to cut samples
- Measure the surface resistance of antistatic flooring and molded products
Note: When used in combination with the DSM-8104 or SM-8220 super megohm meter, Measurement resistance range*: 103 to 1013 Ω
(* When using the SM-8220: 5 × 104 to 1013 Ω)
Dimensions: φ 100mm (3.94in) × 223mm (8.78in), Mass: 2.5 kg (88.2oz) |
|

Measure the surface resistance between the main electrode and ring electrode of the main body electrode. |

Measure the volume resistance of the sample sandwiched between the main electrode and counter-electrode. |
 |
 |
 |
VERIFICATION FIXTURE FOR SURFACE RESISTANCE MEASUREMENT SM9002
The SM9002 Verification Fixture for Surface Resistance Measurement (option) allows you to check the operation of the electrode to increase the reliability of measurement results. |
|
|
|
 |
| Electrode for surface resistance SME-8301 |
 |
 |
Surface resistance can be easily measured by simple pushing the electrode against the specimen. It measures surface resistance of anti-static related goods in combination of mainly SM-8213.
Dimensions: φ 60mm (2.36in) × 50mm (1.97in), Lead length 1m (3.28ft) |
|
|
|
| Electrode for surface resistance SME-8302 |
 |

An electrode distance: 4mm (0.16in) |
Thie is an electrode for surface resistance, when sample is curved shape such as, resin and rubber processed goods, TV cathode tube or sample is small. Surface resistance can be measured by pressing the rubber tips at the tip onto the sample. Electrode interval is 10mm and up to 1010Ω can be measured.
Dimensions: φ 40mm (1.57in) × 115mm (4.53in), Lead length 1m (3.28ft) |
|
|
 |
| Electrode for plate sample SME-8310 |
 |
 |
Sample of 100mm square by up to 8mm in thickness is measurable. The main electrode dia. is 50mm and inner & outer dia. of ring electrode are 70mm & 80mm respectively. Meas. voltage becomes "OFF" while the lid is open to ensure safety. A selector switch allows selection of voltage or surface resistivity.
Dimensions: 215mm (8.46in) W × 78mm (3.07in)H × 165mm (6.50in)D, Lead length 75cm (2.46ft) |
|
|
|
| Electrode for plate SME-8311 |
 |
 |
Sample of 40~100mm square by up to 8mm in thickness is measurable. The main electrode dia. is 19.6mm and inner & outer dia. of ring electrode are 24.1mm & 28.8mm respectively. Meas. voltage becomes "OFF" while the lid is open to ensure safety.
The fundamental specifications are the same as SME-8310.
Dimensions: 215mm (8.46in) W × 78mm (3.07in)H × 165mm (6.50in)D, Lead length 75cm (2.46ft) |
|
|
 |
| Weight electrode SME-8320 |
 |
 |
This is an electrode for plate sample for use together with SME-8350 shield box. This electrode enables extremely easy measurement of surface resistivity and volume of sample with coarse surface such as carpets, etc. The main electrode dia. is 50mm, and the ring electrode inner-dia. and outer-dia. are 70mm and 80mm respectively. |
|
 |
Included: Banana clips ×2
Photo is Combination with Shield box SME-8350 |
|
|
|
| Electrode for liquid sample SME-8330 |
 |

Included: Connection cable Red/ Black each 1, 60cm (1.97ft) |
This is an electrode for liquid sample which is electrically guarded. Total volume is 25ml. Capacitance between main and counter electrode is approx. 45pF. Electrode constant is approx. 500cm. Distance between both electrodes is 1mm. Outer dia. is 36mm, height is approx. 140mm. Resistance up to 1019Ω (at 1000V) can be measured when this electrode is used together with SM-8216.
Note: Included the inspection result sheet
Dimensions: φ 36mm (1.42in) × 140mm (5.51in) |
|
|
 |
| Shield box SME-8350 |
 |

Included rubber sheet |
This is used as a sample accommodation box during measurement of a high-insulation resistance sample, or inductive or capacitive sample to perform electromagnetic shielding.
Dimensions: 250mm (9.84in) W × 100mm (3.94in)H × 200mm (7.87in)D, Lead length 80cm (2.62ft) |
|
|
|
| Electrode for chip capacitor SME-8360 |
 |
 |
For measurement of resistance of tip capacitor, with adjustable jig from 0mm to 11mm. When connected to the meter by interlock cable, meas. voltage becomes "OFF" while the lid is open to ensure safety.
Dimensions: 200mm (7.87in) W × 520mm (2.05in)H × 150mm (5.91in)D, Lead lengt 85cm (2.79ft) |
|
|
 |
| Standard resistor box SR-2 |
 |
 |
This is a resistor box for calibration of the super megohmmeters.
Max. voltage is 1,000VDC and resistor value covers from 10MΩ to 10,000MΩ in 24 points.
Note: Included the inspection result sheet
Dimensions: 270mm (10.63in) W × 90mm (3.54in)H × 195mm (7.68in)D, Lead length 75cm (2.46ft) |
|
|
|
|
 |
|