Leak current measurement, an essential part of electrical safety (For Standard- and Regulation-compliant)
  • LEAK CURRENT HiTESTER ST5541: A low-cost solution that complies with standards governing general-use electrical devices
  • Uninterrupted polarity switching function dramatically reduces cycle time.
  • Support for rated currents up to 20 A gives the instrument more than adequate capability for testing products designed to comply with new standards.
  • Touch panel features simple, interactive operation.
  • Communications functionality and external I/O support allow automatic testing on production lines.
Note: For applications involving leak current measurement of medical-use electrical devices, use the ST5540.
Basic specifications (Accuracy guaranteed for 1 year)
Measurement methods Measurement of voltage drop across body simulated resistance points
Calculation and display of current values True rms measurement
Measurement unit floats relative to instrument ground.
Measurement modes Leak current measurement, voltage measurement, safety conductor current measurement
Standard compliance Electrical Appliances and Materials Safety Act
Measurement of touch current and protective conductor current: IEC 60990 (1999)
Electrical equipment for measurement, control, and laboratory use: IEC 61010-1 (2001)
Information technology equipment: IEC 60950-1 (2005)
Household and similar appliances: IEC 60335-1 (2001) + A1:2004 + A2:2006
Audio, video and similar electronic apparatus: IEC 60065 (2001) + A1:2005
Personnel Protection Systems for EV: UL-2231-1 (2002), UL-2231-2 (2002)
UL: UL-1492 (1996)
Electrical equipment for measurement, control, and laboratory use; current measurement circuits in damp conditions: IEC 61010-1 (2001)
Leak current measurement Ground leak current, 3 types of contact current, free current measurement, 3 types of enclosure leak current
Measurement current DC, AC (true rms, 15 Hz to 1 MHz), AC+DC (true rms, 15 Hz to 1 MHz), AC peak (15 Hz to 1 MHz)
Measurement ranges DC / AC / AC+DC mode:50 μA/500 μA/5 mA/50 mA
AC peak mode:500 μA/1 mA/10 mA/75 mA
Measurement accuracy
(current measurement)
DC measurement: ±2.0% rdg. ±6 dgt. (typ.)
AC / AC+DC measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 100 kHz, typ.)
AC peak measurement: ±2.0% rdg. ±6 dgt. (15 Hz to 10 kHz, typ.)
Interfaces External I/O, USB 1.1 (communications), RS-232C
Functionality Automatic test, data storage for 100 target devices, clock, data backup, printed output (optional), etc.
HiTESTER power supply 100/120/220/240 V AC (specify at time of order), 50/60 Hz, 30 VA rated power
Target device power supply input 100 to 240 V AC, 50/60 Hz Rated current input from terminal block: 20 A
Target device power supply output Output from terminal block: 20 A Output from outlet: 15 A
Dimensions, mass 320 mm (12.60 in) W × 110 mm (4.33 in) H × 253 mm (9.96 in) D, 4.5 kg (158.7 oz)
Supplied accessories TEST LEAD L2200 (Red ×1, Black ×1) ×1 set, ENCLOSURE PROBE 9195 ×1, Instruction manual ×1, Users manual ×1, Power cord ×3, Spare fuse for measurement line ×1
ST5540, ST5541 List of functions
Item ST5540 ST5541
Network Network A
(Electrical Appliances and Materials Safety Act)
Network B (Medical-use electrical devices) -
Network C (IEC 60990)
Network D (UL)
Network E (General-purpose 1)
Network F (General-purpose 2)
Network G (IEC 61010-1)
Major
functions
Power on polarity switching function
Rated current 20 A
Function for checking for blown fuses
Frequency band switching -
110% voltage output terminal (T3 terminal) -
S10, S12, S13, E terminal -
Item ST5540 ST5541
Testing leakage
current mode
Earth leakage current
Touch current
Patient auxiliary current -
Patient leakage current -
Total patient leakage current -
Free current
Enclosure - Earth leakage current
Enclosure - Enclosure leakage current
Enclosure - Line leakage current
Patient leakage current I -
Patient leakage current II -
Patient leakage current III -
OVERVIEW OPTION DOWNLOAD CATALOG