| Data Logger (Temperature, Voltage or other measurement / Multi channels) |
MEMORY HiLOGGER LR8400-20 series |
|
 |
|
| LR8400-20 (with built-in VOLTAGE/TEMP UNIT × 2) |
Built-in units are equivalent to the VOLTAGE/TEMP UNIT LR8500 (15 ch) × 2
Caution: Built-in units cannot be removed or changed |
|
| LR8401-20 (with built-in UNIVERSAL UNIT × 2) |
Built-in units are equivalent to the UNIVERSAL UNIT LR8501 (15 ch) × 2
Caution: Built-in units cannot be removed or changed |
|
| LR8402-20 (with built-in UNIVERSAL UNIT × 1, VOLTAGE/TEMP UNIT × 1) |
Built-in units are equivalent to the UNIVERSAL UNIT LR8501 (15 ch) × 1, and VOLTAGE/TEMP UNIT LR8500 (15 ch) × 1
Caution: Built-in units cannot be removed or changed |
|
 |
 |
|
|
 |
 |
| Portable Data Logger with 30 Standard Channels, Expandible to 60 Channels |
 |
- Compact size despite 30-channel standard capabilities
- Expand up to 30 additional channels
- Protected against unexpected power outages
- Write data to USB memory stick or CF card in real-time
- Built in with USB 2.0 or 100 BASE-TX LAN interfaces
- 5.7" TFT color display
|
| Basic specifications (Accuracy guaranteed for 1 year) |
 |
Analog input UNIVERSAL UNIT LR8501
Actual functionality will depend on combination of units installed
Note: Isolated from each channel to chassis |
[No. of channels] 15 analog channels; isolated scanning method input (4 terminals: push-button type)
[Voltage] 10 mV to 100 V, 1-5 V f.s. 10 ranges, Max. resolution: 500 nV
Note: Isolated between channels and from each channel to chassis
[Temperature: Thermocouples] -200°C to 2000°C (depends on sensor), 3 ranges (K, J, E, T, N, R, S, B, W), Max. resolution 0.01°C
Note: Isolated between channels and from each channel to chassis
[Temperature: Pt 100, JPt 100 sensor] -200°C to 800°C, 3 ranges, Max. resolution 0.01°C
Note: Not isolated between channels
[Resistance] 10 Ω to 200 Ω f.s. 4 ranges, Max. resolution 0.5 mΩ
Note: Not isolated between channels
[Humidity] 100 % rh f.s. 5.0 to 95.0 % rh (when using Z2000), resolution 0.1 % rh
Note: Not isolated between channels nor from each channel to chassis
[Max. rated voltage between isolated input channels] 300 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
|
Analog input VOLTAGE/TEMP UNIT LR8500
Actual functionality will depend on combination of units installed
Note: Isolated from each channel to chassis |
[No. of channels] 15 analog channels; isolated scanning method input (2 terminals: M3 screw type)
[Voltage] 10 mV to 100 V, 1-5 V f.s. 10 ranges, Max. resolution: 500 nV
Note: Isolated between channels and from each channel to chassies
[Temperature: Thermocouples] -200°C to 2000°C (depends on sensor), 3 ranges (K, J, E, T, N, R, S, B, W), Max. resolution 0.01°C
Note: Isolated between channels and from each channel to chassies
[Temperature: Pt 100, JPt 100 sensor] [Resistance] not available
[Humidity] 100 % rh f.s. 5.0 to 95.0 % rh (when using Z2000), resolution 0.1 % rh
Note: Not isolated between channels nor from each channel to chassies
[Max. rated voltage between isolated input channels] 250 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
Caution: Max. voltage from terminals to ground without damage |
| Pulse, Digital input |
[No. of channels] 8 channels, pulse / digital selectable for each channel, M3 screw terminal, not isolated, common ground
[Pulse totalization] 0 to 1000 M pulse, 1 range (No-voltage 'a' contact; normally open, open collector or voltage input), Max. resolution 1 pulse
[Pulse revolutions] 0 to 5000 /n (r/s) f.s. 1 range (same as Pulse totalization input signal condition), resolution 1/n (r/s) Note: "n" is the number of sensor output pulses per revolution, 1 to 1000
[Digital input] Record logical "1" or "0" at each sampling
[Max. rated voltage between input channels] Not isolated
[Max. allowable input] 0 to 50 V |
| Recording intervals |
10 ms to 50 ms, 100 ms to 1 hour, 19 selections (All input channels are scanned at high speed during every recording interval)
Note: limited by using channels at 10 ms to 50 ms interval |
| Digital filter |
Select from OFF/ 50 Hz/ 60 Hz (the cut-off frequency is automatically set) |
| Data storage |
Internal memory: 8 M-words, Data storage media: CF card or USB memory
(Only data recorded to a genuine HIOKI CF card is guaranteed) |
| LAN interface |
100BASE-TX, Functions: Data acquisition using bundled software or PC commands, FTP server, FTP client, HTTP server function, or E-mail system
Note: LAN communication support planned from software Ver. 1.20 |
| USB interface |
USB 2.0 High-speed capable, series mini-B receptacle
Functions: Data acquisition using bundled software or PC commands, Transfer data from the CF card to a PC via USB drive mode
Note: Data transfer not possible from USB memory sticks |
| Display device |
5.7 inch TFT color liquid crystal display (640 × 480 pixel) |
| Other functions |
Save waveform data in real time to the CF card or USB memory stick, Numerical value calculations, Waveform calculations, and others |
| AC Power |
Using the AC ADAPTER 9418-15 (100 to 240 V AC, 50/60 Hz), 7 VA |
| DC Power |
Using the BATTERY PACK Z1000 (optional accessory), Continuous use 5 hr, External power: 10 to 28 V DC (Please contact your HIOKI distributor for connection cord) |
| Dimensions & Mass |
272 mm (10.71 in) W × 182.4 mm (7.18 in) H × 66.5 mm (2.62 in) D, 1.8 kg (63.5 oz), (LR8400-20 main unit, excluding the Battery Pack 370 g/ 13.1 oz) |
| Accessories |
Detailed operating manual ×1, Measurement guide ×1, AC ADAPTER 9418-15 ×1, USB cable ×1, CD-R (data collection software "Logger Utility") ×1 |
|
|