Japanese Chinese Korea
TOA DKK
Super megohm meter SM-8213 < SuperMegOhm Checker & Chart Recorders < Products
SuperMegOhm Checker & Chart Recorders (Electronic Measuring Instruments)

Super megohm meter SM-8213

OVERVIEW OPTION DOWNLOAD CATALOG  
  • Meas. voltages : 5 - 100V DC (6ranges)
  • Meas. range : 0.025 - 2 x 106MΩ (6ranges)
  • Digital/analogue display on LCD
  • Insulation meter for low voltage : 5 - 100V DC
  • Equipped with charge-discharge function (standard spec)
  • Equipped with timer, remote starter, comparator & HV-EN functin (standard spec)
  • Current output as measurement power source : approx. 50mA
  • Compatible for measurement of several sample types withelectrode & other devices
Basic specifications
  SM-8213 SM-8215 SM-8220
Measuring voltage &
Measuring range
5V 2.5x104 to 1x1011Ω    
10V 5x104 to 2x1011Ω   5x104 to 2x1014Ω
15V 7.5x104 to 3x1011Ω    
25V 1.25x105 to 5x1011Ω   1.25x105 to 5x1014Ω
50V 2.5x105 to 1x1012Ω 2.5x105 to 1x1012Ω 2.5x105 to 1x1015Ω
100V 5x105 to 2x1012Ω 5x105 to 2x1012Ω 5x105 to 2x1015Ω
250V   1.25x106 to 5x1012Ω 1.25x106 to 5x1015Ω
500V   2.5x106 to 1x1013Ω 2.5x106 to 1x1016Ω
1,000V   5x106 to 2x1013Ω 5x106 to 2x1016Ω
Accuracy of measuring voltage ± 3% of setting voltage value
Output current Max. 50mA Max. 2mA
Measuring accuracy ± 10% (within 10 times range of min. value on each range at 20°C ), but ± 20% at 108 range of SM-8220
Display LCD (digital & analog display)
Standard function timer, comparator (alarm), remote start, HV-EN, RS-232C
Ambient temperature 0 to 40°C 5 to 35°C
Ambient humidity less than 85%RH
Power Source 100V, 120V, 220V, 240VAC ±10%, but max. 250VAC, 50/60Hz
Power consumption approx.25VA
Dimensions approx.284(w) x 139(h) x 215(d) mm
Weight approx.4.3kgs
Standard accessories: measuring rod(red & black) 1 pair, power code 1pc., instruction manual 1copy
OVERVIEW OPTION DOWNLOAD CATALOG