| SuperMegOhm Checker & Chart Recorders (Electronic Measuring Instruments) |
Digital ultra-insulation / micro ammeter 1-ch DSM-8104 |
|
 |
|
 |
 |
|
|
 |
 |
- Measurement voltage : DC 0.1 - 1,000 V
- Measurement range : 1 x 107 - 3 x 1016 Ω
- Measures insulation resistance of capacitive/highly speed with high accuracy
- High resolution current measurements of 0.1 fA
- Histogram display of selected results
- GP-IB, RS-232, Handler interface equipped as standard - optional LAN interface also available
- Full range of electrodes to suit all types of measurements
|
| Basic specifications |
 |
Measurement range DC Measurement Capability |
 |
| Current Measurement |
| Measurement |
Maximum display |
Resolution |
Accuracy |
| Range Name |
| 10pA |
9.9999pA |
0.1fA |
± (3.0% of rdg+1.2% of range) |
| 100pA |
99.999pA |
1.0fA |
± (1.5% of rdg+0.6% of range) |
| 1nA |
999.99pA |
10fA |
± (0.6% of rdg+0.6% of range) |
| 10nA |
9.9999nA |
100fA |
± (0.4% of rdg+0.5% of range) |
| 100nA |
99.999nA |
1pA |
± (0.4% of rdg+0.5% of range) |
| 1µA |
999.99nA |
10pA |
± (0.4% of rdg+0.5% of range) |
| 10µA |
9.9999µA |
100pA |
± (0.4% of rdg+0.5% of range) |
| 100µA |
999999µA |
1nA |
± (0.4% of rdg+0.5% of range) |
|
|
| Resistance Measurement Capability (1,000V measure voltage) |
 |
| Resistance Measurement |
| Range of Measurement |
Measurement Range Name |
Fundamental Accuracy |
| 1x1014 - 3x1016 (Open-circuit) |
10pA |
± 4.0% of rdg |
| 1x1013 - 1x1014 |
100pA |
± 4.0% of rdg |
| 1x1012 - 1x1013 |
1nA |
± 2.0% of rdg |
| 1x1011 - 1x1012 |
10nA |
± 0.8% of rdg |
| 1x1010 - 1x1011 |
100nA |
± 0.6% of rdg |
| 1x109 - 1x1010 |
1µA |
± 0.6% of rdg |
| 1x108 - 1x109 |
10µA |
± 0.6% of rdg |
| 1x105 - 1x107 |
100µA |
± 0.6% of rdg |
|
|
| Measurement Time Setting |
 |
| Delay |
0 - 9,999ms |
| Sampling Time |
2 - 300ms |
|
|
Voltage Generator Setting Voltage Accuracy and Resolution |
 |
| Setting Voltage Range |
Resolution |
Accuracy |
| 0.1 to 250.0V |
100mV |
± (0.1% of setting + 150mV) |
| 251 to 1,000V |
1V |
± (0.1% of setting + 400mV) |
|
|
| Current Limiter |
 |
| Setting Voltage Range |
Current Limit Value |
| 0.1 to 250.0V |
50mA |
| 10mA |
| 5mA |
| 251 to 1,000V |
10mA |
| 5mA |
|
|
| Measurement Check Function |
 |
| Voltage Monitor, Contact Check Function, |
|
|
| Other Function :Comparator Measurement, Deviation/Percentage Measurement, Surface/Volume Resistivity Measurement |
|
| Interface GP-IB, RS-232C, LAN (Option), Handler Interface |
 |
| Power Supply |
AC100V (115V, 220V, 240V factory option), ± 10%50/60Hz Approx. 55VA |
| Dimension, mass |
Approx. 332W x 89(H x 450D mm Approx. 6.7kg |
| Accessories |
Power cord (1) (Measurement leads are optional.) |
|
|
|