| Component Measuring Instrument (Battery Testers) |
BATTERY HiTESTER BT3563/BT3562 |
|
 |
|
 |
 |
|
|
 |
- Measure high-voltage battery packs up to 300V (BT3563)
- Testing for production line of high-voltage battery pack and battery module
- Large (low-resistance) cell testing
- Choice of PC interfaces for full remote operation
Note: The comparison threshold values depend on the battery manufacturer, type, and capacity, and these must be established by the user. |
 |
- Measurement leads are not included. Purchase the appropriate lead option for your application separately.
- The male (system side) of the EXT I/O connector is also available. Please inquire with your HIOKI distributor.
|
|
| Basic specifications (Accuracy guaranteed for 1 year) |
 |
| |
BT3563/BT3563-01 |
BT3562/BT3562-01 |
| Max. applied measurement voltage |
±300 V DC rated input voltage
±300 V DC maximum rated voltage above ground |
±60 V DC rated input voltage
±70 V DC maximum rated voltage above ground |
| Resistance Measurement ranges |
3 mΩ (Max. display 3.1000 mΩ, resolution 0.1 μΩ) to 3000 Ω (Max. display 3100.0 Ω, resolution 100 mΩ), 7 ranges
Accuracy: 30 mΩ to 3000Ω ranges, ±0.5 % rdg. ±5 dgt. (Add ±3 dgt. for EX.FAST, or ±2 dgt. for FAST and MEDIUM)
3 mΩ range, ±0.5 % rdg. ±10 dgt. (Add ±30 dgt. for EX.FAST, or ±10 dgt. for FAST, or ±5 dgt. for MEDIUM)
Testing source frequency: 1kHz ±0.2Hz, testing current: 100 mA (3 mΩ range) to 10 μA (3000Ω range)
Open-circuit Voltage: 25 V peak (3/30 mΩ ranges), 7 V peak (300 mΩ range), 4 V peak (3Ω to 3000Ω ranges) |
| Voltage Measurement ranges |
DC 6 V (resolution 10 μV) to DC 300 V (resolution 1 mV), 3 ranges |
DC 6 V (resolution 10 μV) to DC 60 V (resolution 100 μV), 2 ranges |
| Accuracy: ±0.01 % rdg. ±3 dgt. (Add ±3 dgt. for EX.FAST, or ±2 dgt. for FAST and MEDIUM) |
| Display |
31000 full digits (resistance), 600000 full digits (voltage), LED |
| Sampling rate |
Four steps
4 ms(Extra-FAST), 12 ms(FAST), 35 ms(Medium), 150 ms(Slow)
(Typ., sampling time depends on supply frequency settings and function.) |
| Measurement time |
Response time + sampling rate, approx. 10 ms for measurements
(Response time depends on reference values and the measurement object.) |
| Comparator functions |
Setting: Upper and lower limit
Output: 3 levels (Hi, In, Lo) or (Pass, Fail), Open-collector(35V, DC50mA max), Display, Dual audible indicator |
| Analog Output |
Measured resistance (displayed value, from 0 to 3.1V DC, -01 suffix models only) |
| Interface |
External I/O, RS-232C, Printer (RS-232C), GP-IB(-01 suffix models only) |
| Power supply |
100 to 240 V AC, 50/60 Hz, 30 VA max. |
| Dimensions, mass |
215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.4 kg (84.7 oz) |
| Supplied accessories |
Power cord × 1 |
|
|