OVERVIEW  
Basic specifications
Measurement functions Number of channels: 2, for distortion measurement (electronic auto-balancing, balance adjustment range within ±10000 με)
Input connectors Via adapter cable, TAJIMI PRC03-12A10-7M10.5, Max. rated voltage to earth:
33 Vrms or 70 V DC (with input isolated from the unit, the maximum voltage that can be
applied between input channel and chassis and between input channels without damage)
Suitable transducer Strain gauge converter, bridge impedance: 120 Ω to 1 kΩ (bridge voltage 2 V),
350 Ω to 1 kΩ (bridge voltage 5 V, 10 V), bridge voltage 2, 5, 10 ±0.05 V
Measurement range 20 με - 1000 με/division, 6 ranges, full scale: 20 divisions, lowpass
filter: 5/10/100/1 kHz
Anti-aliasing filter Integrated filter for suppressing aliasing distortion caused by FFT
processing (automatic cutoff frequency setting/OFF)
Measurement resolution 1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860 series)
Highest sampling rate 200 kS/s (2-channel simultaneous sampling)
Accuracy After auto-balancing DC amplitude: ±(0.4% of full scale +2 με), zero position: ±0.1% of full scale
(at 5 Hz filter ON)
Frequency characteristics DC to 20 kHz +1/-3 dB
Max. allowable input 10 V DC (the maximum voltage that can be applied across input pins without damage)