| Measurement functions |
Number of channels: 1, voltage measurement |
| Input |
Isolated BNC connector, Max. rated voltage to earth: 370 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channels and chassis, or across input pins, without causing damage) |
| Measurement range |
5 mV to 20 V/division, 12 ranges, full-scale = 20 divisions, AC voltage for possible measurement / display using the memory function: 280 V rms, low-pass filter: 5/50/500/5 k/50 kHz |
| Measurement resolution |
Data in the 1/1600 measurement range (using 16-bit A/D) |
| Maximum sampling rate |
1 MS/s |
| Accuracy |
DC amplitude: ±0.2% f.s. Zero-position: ±0.1% f.s. |
| Frequency characteristics |
DC to 100 kHz ±3 dB. With AC coupling: 7 Hz to 100 kHz ±3 dB |
| Input resistance and capacitance |
1 M-ohm, 40 pF (when C is 100 kHz) |
| Input coupling |
DC, GND, AC |
| Max. allowable input |
400 V DC (the maximum voltage that can be applied between input pins without causing damage) |
| Power terminal |
DIFFERENTIAL PROBE 9322 Power (the POWER CORD 9328 is required for connection) |
| Anti-aliasing filter |
Cutoff frequency (fc): 20 Hz to 40 kHz (set automatically) Attenuation: -66 dB or greater at 1.5 fc |
| Accessories |
None * The input cord and conversion cable are optional |