Japanese Chinese Korea
VOLTAGE/CURRENT UNIT 8951 < Recorder < Products
Recorder Option (Input units for the MEMORY HiCORDER series)

VOLTAGE/CURRENT UNIT 8951

OVERVIEW OPTION  
Voltage and Current input Amplifier
This unit is one input channel. Simply slot the 8951 into the Digital Oscillo Series MEMORY HiCORDER 8855 for use. In addition to voltage measurement, this single unit handles current measurement.
Basic specifications
Measurement functions Number of channels: 1, voltage measurement/current measurement using a clamp
Input Metal BNC connector, Max. rated voltage to earth: 30 V RMS or 60 V DC (with input isolated from the unit, the maximum voltage that can be applied between input channels andchassis, or across input pins, without causing damage, ±12 V common ground for installed units when using a clamp)
Sensor connector: CLAMP SENSOR 9270s input (±12 V common ground for installed units)
Voltage measurement range 1 mV to 5 V/division, 12 ranges, full-scale = 20 divisions, AC voltage for possible measurement / display using the memory function: 30 V rms, low-pass filter: 5/500/100 k/1 MHz
Current measurement range With the 9270, 9272-10 (20 A), 9277, 3273, 3273-50 or 3276: 10 mA to 5 A/division, 9 ranges, full-scale = 20 divisions
With the 9271, 9272-10 (200 A), 9278, or 3274: 100 mA to 50 A/division, 9 ranges, full-scale = 20 divisions
With the 9279: 200 mA to 100 A/division, 9 ranges, full-scale = 20 divisions
With the 3275: 100 mA to 100 A/division, 10 ranges, full-scale = 20 divisions
Low-pass filter: 5/500/100 k/1 MHz
(1 MHz: 3273/3274/3275, 100 kHz: When using the 3273, 3274, 3275, 9277, or 9278)
Measurement resolution Data in the 1/100 measurement range (using 12-bit A/D)
(With the 9279: part of the current range or in the 1/80 measurement range)
Maximum sampling rate 20 MS/s
Accuracy DC amplitude: ±0.5% f.s. Zero-position: ±0.15% f.s.
(add the accuracy and characteristics of the sensor or probe used when measuring current)
Frequency characteristics DC to 4 MHz ±3 dB, With AC coupling: 7 Hz to 4 MHz ±3 dB
Input resistance and capacitance 1 M-ohm, 50 pF (when C is 100 kHz)
Input coupling DC, GND, AC
Max. allowable input 30 V rms or 60 V DC (the maximum voltage that can be applied between input pins without causing damage)
Power terminal ±12 V for the 3273/3273-50/3274/3275
(common ground with the power terminals of installed units)
Accessories None * The input cord and conversion cable are optional
OVERVIEW OPTION