| Measurement functions |
Number of channels: 1, voltage measurement |
| Input |
Isolated BNC connector, Max. rated voltage to earth:370 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channels and chassis, or across input pins, without causing damage) |
| Measurement ranges |
5 mV to 20 V/division, 12 ranges, full-scale = 20 divisions, AC voltage for possible measurement and display using the memory function: 280 V RMS Low-pass filter: 5/500/5 k/1 MHz |
| Measurement resolution |
Data in the 1/100 measurement range (using 12-bit A/D) |
| Maximum sampling rate |
20 MS/s |
| Accuracy |
DC amplitude: ±0.4 %f.s. Zero-position: ±0.1 %f.s. |
| Frequency characteristics |
DC to 10 MHz ±3 dB with AC coupling: 7 Hz to 10 MHz ±3 dB |
| Max. allowable input |
400 V DC (the maximum voltage that can be applied between input pins without causing damage) |
| Power terminal |
DIFFERENTIAL PROBE 9322 Power (the POWER CORD 9328 is required for connection) |
| Accessories |
None * The input cord is optional |