Version Upgrade History for MEMORY HiCORDER MR8847-01, MR8847-02, MR8847-03 and 8847

Firmware Version (Storage FPGA)
Update time Version number Content
Oct. 2011 V3.03 Added functions
1. Added [Hex] to [Format] item.
Resolved bugs
1. When you change the [Format] setting of X-Y recorder function, its printing format cannot be changed.(Applies only to units with firmware version 3.02)
2. Other minor bugs were resolved.
Jun. 2011 V3.02 Added functions
1. The Roll mode function for X-Y waveform.
Resolved bugs
1. Problems with auto-saving data or auto-printing data while using the memory division feature.
2. Other minor bugs were resolved.
Feb. 2011 V2.07 Added functions
1. Chinese display screen
Dec. 2010 V2.06 Resolved bugs
1. Fixed an error in regard to end code of BMP compressed file.
2. Other minor bugs were resolved.
Jun. 2010 V2.05 Resolved bugs
1. When the speed of LAN communication is in 10BASE(or the speed of LAN communication is low), during FTP file transfer, communication may stop and some of the transferred data may be corrupted (MAC address and measurement data).
Mar. 2010 V2.04 Added functions
1. The batch copy function to copy Unit related settings.
2. Now you can select from two file sizes to save Waveform image data.
Resolved bugs
1. When displaying the “XYSingle” or “XYQuad” format for the Memory function, if you printed or saved the waveform data after stopping measurement, there were cases when a non-existent data extra was included.
2. Other minor bugs were resolved.
Feb. 2010 V2.03 Resolved bugs
1. When copying the scaling setting values to another channel on the scaling screen, if the variable function is enabled on the copy destination channel, the upper and lower limits can not be set correctly, so that the size of the waveform is not displayed correctly.
2. Other minor bugs were resolved.
Jan. 2010 V2.02 Resolved bugs
1. When performing a time difference calculation or a phase contrast calculation, if the channel of the phenomenon A was set by the waveform calculation number and the channel of the phenomenon B was set as analog channel, the level value for the phenomenon B was automatically replaced by the level value for the phenomenon A.
2. Other minor bugs were resolved.
Aug. 2009 V2.01 Added functions
1. FFT Function
2. Memory Division Function
3. Waveform Calculation Function
4. 8970/8971units are now applicable
5. Some numerical calculations were added: Time difference calculation, Phase contrast calculation, High/Low level calculation
6. Trigger search function
Resolved bugs
1. When the units configuration of the loaded file was different from the currently installed units configuration, the setting file was found not to be loaded correctly
2. When performing Pen up/down operation 1000 times or more and then printing waveforms with line interpolation function, the portion that should not be interpolated was found to be interpolated.
3. Other minor bugs were resolved.
Feb. 2009 V1.06 Improvement
1. The multiple makers' internal memory chips are now applicable.
* This version upgrade will not affect the performance of your 8847.
Dec. 2008 V1.05 Resolved bug
1. When the waveforms of 9 channels or more was printed manually at a print speed set to "Fast", the printing was found to stop halfway.
Dec. 2008 V1.04 Resolved bug
1. When turning on the power, checksome error for the unit not being installed was found to appear rarely.
Nov. 2008 V1.03 Resolved bug
1. When a trigger event occurred by Standard logic signal if the waveform changed faster than the sampling interval, the waveform change at the point of triggering was found not to be displayed on the screen.
2. Other minor bugs were resolved.
Oct. 2008 V1.02 Added function
1. Korean display screen
Resolved bugs
1. When the units configuration of the loaded file was different from the currently installed units configuration, the waveform data was found not to be displayed correctly on the screen.
2. Other minor bugs were resolved.
FPGA (IO LCDC)
Update time Version number Content
Mar. 2010 IO:0103
LCDC:0103
Improvement
1. The version upgrade program has been changed, because upgrade was rarely inevitable.
2. During upgrading, its progress can be seen on the screen now.
* No functional change.
Aug. 2009 IO:0103
LCDC:0103
Improvement
1. Change of internal circuit
* This version upgrade will not affect the performance of your 8847.
Jun. 2008 IO:0101
LCDC:0103
Resolved bugs
1. There was a case when display screen vanished.