Japanese Chinese Korea
MEMORY HiLOGGER 8430-20 < Data Logger < Products
Data Logger (Temperature, Voltage or other measurement / Multi channels)

MEMORY HiLOGGER 8430-20

CE
compliant
USB2.0
OVERVIEW OPTION DOWNLOAD CATALOG VERSION UPGRADE  
Small and light enough for the palm of your hand !
  • 10 msec maximum sampling period simultaneous all channels (rapid scanning type)
  • Tough against noise with strengthened input hardware
  • Large and beautiful display (4.3 inch TFT liquid crystal display)
  • Plus 4 pulse-counting inputs
  • CompactFlash card makes direct recording a snap
Note: The 8430-20 is not bundled with the Battery Pack 9780. Use only PC cards sold by HIOKI.
Thermocouples are not provided by HIOKI, and must be purchased from a separate vendor.
Basic specifications (Accuracy guaranteed for 1 year)
No. of channels Analog 10 channels isolated scanning method input. (M3 mm dia. screw terninal block)
Pulse input 4 channels (All pulse inputs share common ground with the main unit)
Measurement parameters Voltage : ±100 mV to ±60 V, 1-5V f.s. 6 ranges, Max. resolution 5μV
Temperature (thermocouples) : -200°C to 2000°C (depend on the sensor), 1 range (K, J, E, T, N, R, S, B), Max. resolution 0.1°C
Temperature (Pt 100 sensor) : not available
Humidity : not available
Totalized pulses : 0 to 1000M count, 1 range (No-voltage 'a' contact, open collector or voltage input), Max. resolution 1 count
Rotation count : 0 to 5000/n (r/s) f.s. 1 range (No-voltage 'a' contact, open collector or voltage input), Resolution 1/n (r/s) Note: n = pulses per rotation (1 to 1,000)
Max. allowable input DC 60V (Analog input), DC -5V to 10V (Pulse input)
Max. rated voltage to earth DC 60V, AC 30Vrms (Upper limit voltage that does not cause damage when applied between input channel and chassis, and between each input channels)
Recording intervals 10ms to 1hour, 19selections (All input channels are scanned at high speed during every recording interval)
Selectable Filters 50Hz, 60Hz, or OFF (digital filtering of high frequencies on analog channels)
Data Recording Capacity Internal storage: 3.5MWords (7MB of two-byte data points, or four-byte pulse measurements)
External storage: Up to 2GB (HIOKI CF cards only)
External Interface One USB 2.0 series mini B receptacle
Functions:Control from a PC(Ver 1.00 or later),Transfers internal data on the CF card to a PC(Ver 1.10 or later)
Display 4.3 inch WQVGA-TFT color LCD (480 × 272 dots)
Function Save data to the CF Card in real time, Numerical Calculation, External trigger input, Trigger output, Alarm output, Scaling
Power Sources (1) 100 to 240V AC, 50/60Hz using AC Adapter Model 9786
(2) Battery Pack Model 9780, Continuous operating time: Approx. 2.5 hours
(3) 12V battery (10 to 16V DC ±10%, please contact HIOKI for connection cord)
Dimensions & Mass 176mm(6.93in) W × 101mm(3.98in) H × 41mm(1.61in) D, 550g(19.4oz) (HiLOGGER only)
Accessories Instruction Manual × 1, Measurement Guide × 1, Application Disk (Logger Utility program) × 1, USB cable × 1, AC Adapter 9786 × 1, Shoulder Strap × 1, Protection Sheet 9809 × 1
OVERVIEW OPTION DOWNLOAD CATALOG VERSION UPGRADE