| Data Logger (Temperature, Humidity or other measurement / Multi channels) |
MEMORY HiLOGGER 8420-51 |
|
 |
|
 |
 |
|
|
 |
 |
All analog inputs fully isolated! Voltage/temperature loggers with PC networking support |
 |
- 8420-51(8ch), 8421-51(16ch), 8422-51(32ch) All analog channels isolated.
- Temperature, voltage, humidity, cumulative pulses, rotations
- Max. 32-channel, 100ms high-speed multi-channel data collection
- Simultaneous numerical and graphical display
- Attachable Printer and I/O Module
- Up to 16 HiLOGGERs can be connected and controlled via LAN.
|
 |
| Basic specifications |
 |
| Measurement items |
Temperature, voltage, humidity (with optional sensor), cumulative pulses, rotations |
| Input channels |
Analog 8 channels isolated PHOTO-MOS relays Pulse Inputs: 4 ch, Logic Inputs: 16 ch (using the 8993 DIGITAL I/O UNIT) |
| Thermocouple types |
K, E, J, T, N, W (WRe5-26), R, S, and B |
| Platinum measurement resistance types |
Pt 100, JPt 100, 3- and 4-line types |
| Analog voltage input |
5 ranges: 100 mV, 1 V, 10 V, 100 V and 1-5 V f.s. |
| Recording interval |
100 ms to 1 hour (5 s to 1 hour when measuringhumidity). Note: All input channels are scanned at high speed during each recording interval. |
| Digital Filter |
OFF/50Hz/60Hz (when set to 50 or 60Hz), the appropriate digital filter is automatically set according to the recording interval |
| Memory capacity |
Internal: 32 MB (16MW data points), External:up to 1G (9726/ 9727/ 9728/ 9729 PC card, 128MB to 1GB) |
| Real-Time save |
Waveforms are saved as binary data to the PC Card in real time. Stored data can be recalled by the instrument in blocks of 32 MB each. |
| PC interface |
RS-232C and 10Base-T Ethernet connectors |
Interfaces (optional, sold separately) |
8992 PRINTER UNIT (Detachable type) or 8993 DIGITAL I/O UNIT (Detachable type) can be connected |
| Recording and display |
5.7-inch STN color LCD, 112 mm (4.4 in) × 18 m (59.06 ft), thermal paper roll |
| Power supply |
(1) AC ADAPTER 9418-15 (12 V DC ±5 %), (2) BATTERY PACK 9447 (NiMH, when used with AC adapter, the adapter has priority), (3) 12V car battery (using connection cable – please inquire) |
| Dimensions, mass |
234 mm(9.21 in)W × 170 mm(6.69 in)H × 52 mm(2.05 in)D, 1.4 kg (49.39 oz) (instrument only, without battery) 310 mm(12.20 in)W × 170 mm(6.69 in)H × 52 mm(2.05 in)D, 1.7 kg (59.98 oz) (with printer, without battery) 302 mm(11.89 in)W × 170 mm(6.69 in)H × 52 mm(2.05 in)D, 1.7 kg (59.98 oz) (with Digital I/O Unit, without battery) |
| Accessories |
AC ADAPTER 9418-15 (1), Terminal Cover (1), Screwdriver (1), Wave viewer software (1), Interface Manual FD (1) |
|
|