| Generator functions and Accuracy |
Constant voltage: 0 to ±2.5 V, 100 µV resolution, 0 to ±25 V, 1 mV resolution, Sink /source: ±25 mA, ±0.03 % of setting ±3 dgt.
Constant current: 0 to ±25 mA, 1 µA resolution, Sink /source; ±25 V, ±0.03 % of setting ±3 dgt. |
| Thermoelectric power |
A TC(0 °C)/TC(RJ) K: -176.0 to 1372.0 °C, E: -220.0 to 839.0 °C, J: -208.0 to 1108.0 °C, T: -169.0 to 400.0 °C, R: -50 to 100 °C, 101 to 1769 °C, S: -50 to 100 °C, 101 to 1769 °C, B: 300 to 600 °C, 601 to 1820 °C, |
| Basic Accuracy |
±0.05 % of setting ±0.5 °C (at Thermoelectric power K) |
| Measurement functions and Accuracy |
Voltage: 0 to ±2.5 V, 100 µV resolution, 0 to ±25 V, 1 mV resolution, Input resistance: 1 m-ohm, ±0.03 % rdg. ±2 dgt. Current: 0 to ±25 mA, 1 µA resolution, Input resistance: 25 Ohm, ±0.03 % rdg. ±2 dgt. |
| Temperature |
-25.0 to 80.0 °C, 0.1 °C resolution (use the 9184) |
| Standard resistance |
100 Ohm, ±0.2 % rdg. |
| Power supply |
LR6(alkaline) 6 pieces, or 9418-10 AC ADAPTER |
| Dimensions, mass |
104 mm(4.09 in)W × 180 mm(7.09 in)H × 58 mm(2.28 in)D, 590 g (20.8 oz) |
| Accessories |
INPUT CORD 9168 (1), TEST LEAD 9170 (1), Fuse(3) |
|