| Safety Standards Measuring Instrument (AC Withstand Voltage Testers) |
HIGH VOLTAGE SCANNER 3930 |
|
 |
|
 |
 |
|
|
 |
 |
| Automatic Multipoint Testing of High Voltages |
 |
- Maximum of 32 Channels
In single mode, an individual 3930 can measure eight channels, and up to four units can be connected together for a total of 32 channels.In multi-mode, the 3930 has four high and low channels, respectively;and when four units are connected, up to 16 high and low channels can be tested.
- Safe Design
The 3930 can handle up to 5kV of direct current or alternating current.
- The 3930 is a general-purpose high voltage scanner that can not only be controlled
by using the HIOKI 3153's program function, but also by general-purpose logic and sequencers.
- Primary Applications
Insulation withstand voltage testing for electrical and electronic equipment
Insulation withstand voltage testing for electric components |
 |
| Functions |
 |
| Operation modes |
Multi- and single modes ( External switch ) |
| Number of channels |
Multi-mode; 4 high channels and 4 low channels Single mode; 8 high channels and a common channel |
| Rated voltage used |
AC 5 kV/DC 5 kV |
| Control method |
General-purpose control |
|
|
| Relay area |
 |
| Maximum open and closed voltage |
5000 V DC, 5000 V AC |
| Maximum open and closed current |
1.0 A (open and closed capacity: 50 W) |
| Contact point indirect contact resistance |
500 mor less, with 1 mA AC |
| Contact point maximum capacity |
50 W |
| Operation time |
6 ms or less |
| Recovery time |
6 ms or less |
|
|
| General specifications |
 |
| Degree of Accuracy |
Standards for current leakage when applying voltage
Single mode, no output cable, and all output relays Turned on for both AC and DC.
When applying DC (1000 V); 0.1 µA or less/unit When applying AC (5 kV, 50/60 Hz only); 0.4 mA or less/unit
(Differs depending on the status of the connection cable) |
| Operation temperature range |
0°C to 40°C, 80% rh or less (no condensation) |
| Storage temperature range |
-10°C to 50°C, 90% rh or less (no condensation) |
| Operation environment |
Indoors, altitude of 2000 m or less |
| Withstand voltage |
High voltage terminal - between the chassis: AC 10 kV, 10 mA, 1 min |
| Power |
VSCV 24 V DC, ±10% (applied using the control signal input connector) |
| Maximum rated power |
12 VA |
| Measurements |
Approx. 316 (W) × 100 (H) × 350 (D) mm |
| Mass |
Approx. 4.2 kg |
| Standard accessories |
Connection cables H.V. TEST LEAD 9615 |
|
|