Japanese Chinese Korea
BATTERY HiTESTER 3561, 3561-01 < Component Measuring < Products
Component Measuring Instrument (Battery Testers)

BATTERY HiTESTER 3561, 3561-01

CE
marked
GP-IB
3561-01
RS-232C
OVERVIEW OPTION DOWNLOAD CATALOG EXTRAS  
For high-speed production line testing of small battery packs
  • High-speed testing for production line of small battery packs for mobile and portable communications devices
  • Measure internal resistance and battery voltage
  • For process control such as in high-speed automated assembly lines.
Note: The comparison threshold values depend on the battery manufacturer, type, and capacity, and these must be established by the user.
  • Measurement leads are not included. Purchase the appropriate lead option for your application separately.
  • The male (system side) of the EXT I/O connector is also available. Please inquire with your HIOKI distributor.
Basic specifications (Accuracy guaranteed for 1 year)
Max. applied measurement voltage ±22 V DC
±70 V DC maximum rated voltage above ground
Resistance Measurement ranges 300 mΩ (Max. display 310.00 mΩ, resolution 10 μΩ) to 3 Ω (Max. display 3.1000 Ω, resolution 100 μΩ), 2 ranges
Accuracy: ±0.5 % rdg. ±5 dgt. (Add ±3 dgt. for EX.FAST, or ±2 dgt. for FAST and MEDIUM)
Testing source frequency: 1 kHz ±0.2 Hz, testing current: 10 mA (300 mΩ range), 1 mA (3 Ω range)
Open-circut Voltage: 7 V peak7
Voltage Measurement ranges DC 20 V, resolution 0.1 mV
Accuracy: ±0.01 % rdg. ±3 dgt. (Add ±3 dgt. for EX.FAST, or ±2 dgt. for FAST and MEDIUM)
Display 31000 full digits (resistance), 199999 full digits (voltage), LED
Sampling rate
Four steps:   4 ms(Extra-FAST), 12 ms(FAST),
35 ms(Medium), 150 ms(Slow)
(Typ., sampling time depends on supply frequency settings and function.)
Measurement time Response time + sampling rate, approx. 3 ms for measurements
(Response time depends on reference values and the measurement object.)
Comparator functions Dual comparators independently judge Hi/IN/Lo states of resistance and voltage
Setting: Upper and lower limit, Deviation (%) from a standard value
Output: 3 levels (Hi, In, Lo), besides independent judgment results for resistance and voltage, ANDed calculation results
Display, Dual audible indicato, or external I/O terminal, Open-collector (35 V, 50 mA DC max.)
Interface External I/O, RS-232C, Printer (The 9670 via RS-232C), GP-IB (-01 suffix models only)
Power supply 100 to 240 V AC, 50/60 Hz, 30 VA max.
Dimensions, mass 215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.4 kg (84.7 oz)
Supplied accessories Power cord ×1