Japanese Chinese Korea
RESISTANCE HiTESTER 3541 < Component Measuring < Products
Component Measuring Instrument (For low resistance measurement)

RESISTANCE HiTESTER 3541

CE GP-IB RS-232C
OVERVIEW OPTION DOWNLOAD CATALOG  
  • Wide Measurement Range
    0.1 µΩ(20 mΩ range) to 110.000 MΩ
  • High Speed & High Precision Measurements
    As fast as 0.6 ms with 70 ppm precision (in the 2 kΩ to 110 kΩ range)
  • Two Types of Temperature Correction
    Correction by Pt sensor or infrared thermometer
  • Equipped with EXT I/O, GP-IB and RS-232C interfaces
    Easily integrates into automated production lines
Basic specifications (Accuracy guaranteed for 1 year)
Resistance range 20 mΩ range (Max. 20.0000 mΩ, 0.1 μΩ resolution) to 100 MΩ range (Max. 110.000 MΩ, 1 kΩ resolution), 11 step
[at Low Power ON] 2 Ω range (Max. 2.00000 Ω, 10 μΩ resolution) to 2 kΩ range (Max. 2.00000 kΩ, 10 mΩ resolution), 4 steps
Measurement accuracy (SLOW 2) ±0.007 % rdg. ±0.0015 % f.s. dgt. (at 2 k/ 20 kΩ range)
[at Low Power ON] ±0.011 % rdg. ±0.01 % f.s. (at 2 to 2 kΩ range)
Testing current 100 nA to 1 A DC, [at Low Power ON] 10 μA to 10 mA DC
Open-terminal voltage 5 V DC (at 20/ 200mΩ range), 2.6 V DC (at 2 to 20kΩ range), 13 V DC (at 1M to 100MΩ range), [at Low Power ON] 60 mV DC
Sampling rate
Delay = 0ms, TC off,
Statistics operation off,
Offset voltage revise off
FAST MIDIUM SLOW 1 SLOW 2
50Hz 0.6ms ± 0.3ms 21ms ± 1ms 155ms ± 5ms 455ms ± 10ms
60Hz 0.6ms ± 0.3ms 17ms ± 1ms 149ms ± 5ms 449ms ± 10ms
Response time (at truely resistance) : the above time plus approx. 1 ms
(different at setting condition, or testing target)
Other functions Temperature measurement (-10.0 to 99.9 °C), Revised resistance with temperature, Revised temperature with inductor resistance, Comparator, Settings save & load (Max. 30 settings), Classification measurement (BIN function), Statistics operation, Offset voltage revise, Delay, Average function
Interface RS-232C and GP-IB (standard)
External I/O [Output] BCD, End of measurement, Comparator result, BIN result, NG, with open collector output
[Input] Select setting, Trigger, Calibration, Zero-adjust, Print, with C-MOS level
Power supply, consumption 100 to 240 V AC, 50/60 Hz, 30 VA max.
Dimensions, mass 215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.6 kg (91.7 oz)
Supplied accessories CLIP TYPE LEAD 9287-10 × 1, TEMPERATURE PROBE 9451 × 1, Power cord × 1, EXT I/O Male connector × 1, Operation manual × 1
OVERVIEW OPTION DOWNLOAD CATALOG