| Component Measuring Instrument (For low resistance measurement) |
RESISTANCE HiTESTER 3541 |
|
 |
|
 |
 |
|
|
 |
 |
| Wide measurement range & high-resolution, Optimized for Automated Systems |
 |
- Wide Measurement Range, 0.1 μΩ(20 mΩ range) to 110.000 MΩ
- High Speed & High Precision Measurements, As fast as 0.6 ms with 70 ppm precision (in the 2 kΩ to 110 kΩ range)
- Two Types of Temperature Correction, Correction by Pt sensor or infrared thermometer
- Equipped with EXT I/O, GP-IB and RS-232C interfaces, Easily integrates into automated production lines
|
 |
| The CLIP TYPE LEAD 9287-10 is bundled. As far as concerned other test fixtures, purchase the appropriate lead option for your application separately. |
| Basic specifications (Accuracy guaranteed for 1 year) |
 |
| Resistance range |
20 mΩ range (Max. 20.0000 mΩ, 0.1 μΩ resolution) to 100 MΩ range (Max. 110.000 MΩ, 1 kΩ resolution), 11 steps
[at Low Power ON] 2 Ω range (Max. 2.00000 Ω, 10 μΩ resolution) to 2 kΩ range (Max. 2.00000 kΩ, 10 mΩ resolution), 4 steps |
| Measurement
accuracy (SLOW 2) |
±0.007 % rdg. ±0.0015 % f.s. dgt. (at 2 k/ 20 kΩ range)
[at Low Power ON] ±0.011 % rdg. ±0.01 % f.s. (at 2 to 2 kΩ range) |
| Testing current |
100 nA to 1 A DC, [at Low Power ON] 10 μA to 10 mA DC |
| Open-terminal voltage |
5 V DC (at 20/ 200mΩ range), 2.6 V DC (at 2 to 20kΩ range), 13 V DC (at 1M to 100MΩ range), [at Low Power ON] 60 mV DC |
Sampling rate
Delay = 0ms, TC off,
Statistics operation off,
Offset voltage revise off |
|
FAST |
MIDIUM |
SLOW 1 |
SLOW 2 |
| 50Hz |
0.6ms ± 0.3ms |
21ms ± 1ms |
155ms ± 5ms |
455ms ± 10ms |
| 60Hz |
0.6ms ± 0.3ms |
17ms ± 1ms |
149ms ± 5ms |
449ms ± 10ms |
Response time (at truely resistance) : the above time plus approx. 1 ms
(different at setting condition, or testing target) |
|
|
| Other functions |
Temperature measurement (-10.0 to 99.9°C), Revised resistance with temperature, Revised temperature with inductor resistance, Comparator, Settings save & load (Max. 30 settings), Classification measurement (BIN function), Statistics operation, Offset voltage revise, Delay, Average function |
| Interface |
RS-232C and GP-IB (standard) |
| External I/O |
[Output] BCD, End of measurement, Comparator result, BIN result, NG, with open collector output
[Input] Select setting, Trigger, Calibration, Zero-adjust, Print, with C-MOS level |
| Power supply, consumption |
100 to 240 V AC, 50/60 Hz, 30 VA max. |
| Dimensions, mass |
215 mm (8.46 in) W × 80 mm (3.15 in) H × 295 mm (11.61 in) D, 2.6 kg (91.7 oz) |
| Supplied accessories |
CLIP TYPE LEAD 9287-10 × 1, TEMPERATURE PROBE 9451 × 1, Power cord × 1, EXT I/O Male connector × 1, Operation manual × 1 |
|
|