LCR HiTESTER 3532-50 < Component Measuring < Products
Component Measuring Instrument (LCR Testers for Measuring Inductance, Capacitance, or Impedance)

LCR HiTESTER 3532-50

CE GP-IB
OPTION
RS-232C
OPTION
OVERVIEW OPTION DOWNLOAD CATALOG EXTRAS  
Increased measurement speed of up to 5msec !
42Hz to 5MHz, wide test frequency range LCR meter
  • High speed measurement of 5ms
  • High precision measurement of ±0.08% basic accuracy
  • Interactive touch panel operation
  • Zoom feature for easy viewing
  • Print measurement values and comparator results (option)
This product is not supplied with measurement probes or test fixtures.
Please select and purchase the measurement probe or test fixture options appropriate for your application separately.
Basic specifications (Accuracy guaranteed for 6 months)
Measurement parameters |Z|,|Y|, θ, Rp, Rs(ESR), G, X, B, Cp, Cs, Lp, Ls, D(tan δ), Q
Measurement ranges 100 mΩ to 100 MΩ, 10 ranges (All parameters defined by |Z|)
Extent of Measurement Impedance |Z|, R, X: 10.00 mΩ to 200.00 MΩ (depending on condition)
θ: -180.00° to +180.00°, C: 0.3200 pF to 370.00 mF, L: 16.000 nH to 750.00 kH, D: 0.00001 to 9.99999, Q: 0.01 to 999.99, |Y|, G, B: 5.0000 nS to 99.999 S (Note: All measurement ranges except for |Z| are for reference only)
Basic accuracy |Z|: ±0.08% rdg., θ: ±0.05°
Source frequency 42Hz to 5MHz (0.1Hz to 1kHz steps)
Measurement signal level 10mV to 5V rms (up to 1MHz), 50mV to 1V rms (1MHz to 5MHz), (1mV rms steps) 10μA to 100mA rms (up to 1MHz), 50μA to 20mA rms (1MHz to 5MHz), (10μA rms steps)
Output resistance 50Ω
Display LCD with backlight display, 99999 (3, 4, or 5 digits unit setting possible)
Measurement times FAST: 5ms, NORMAL: 21ms,
SLOW 1: 72ms, SLOW 2: 140 ms (typical values for displaying |Z|)
DC bias Superimposed DC voltage, DC current to source signal (used with the optional DC bias unit and constant voltage or current source equipment)
Function Comparator, External input/Output (EXT. I/O), GP-IB or RS-232C interface (option)
(Note: RS-232C interface required if used with the Printer 9442.)
Power consumption Selectable 100, 120, 220 or 240V AC ±10%, 50/60Hz, 50VA max.
Dimensions, mass 352mm (13.86in) W × 124mm (4.88in) H × 323mm (12.72in) D, 6.5kg (229.3oz)
Supplied accessories Power cord × 1, Spare fuse × 1