| Power Measuring Instruments (For DC or AC line Power measurement) |
AC/DC POWER HiTESTER 3334, 3334-01 |
|
 |
|
AC/DC POWER HiTESTER 3334
AC/DC POWER HiTESTER 3334-01 (GP-IB installed model) |
|
 |
 |
|
|
 |
 |
| Solves All of your Energy Consumption Testing Needs |
 |
- DC measurement mode, AC, and AC+DC measurement possible
- Integration function for current and power
- High basic accuracy ±0.2%
- Extended Period of Guaranteed Accuracy of 3 Years
- Complete Accuracy Over a Wide Input Range
1.00mA〜 30.00A, 0.150V〜 300.0V, 0.0000W〜 9.000kW |
 |
| Basic specifications (Accuracy guaranteed for 3years) |
 |
| Measurement lines |
Single-phase/ two-wires |
| Measurement items |
Voltage, Current, Active power, Apparent power, Power factor, Frequency, Integration (current, active power), Waveform peak (voltage and current) |
| Measurement ranges |
[Voltage] AC/DC 15.000/30.00/150.00/300.0V
[Current] AC/DC 100.00/300.0 mA, 1.0000/3.000/10.000/30.00A
[Power] 1.5000 W〜9.000 kW (combination of voltage and current ranges) |
Integration measurement
Integration time up to 10,000 hours |
[Current] No.of displayed digits: 6 digits (from 0.00000mAh, Polarity-independent integration and Sum value)
[Active power] No.of displayed digits: 6 digits (from 0.00000mWh, Polarity-independent integration and Sum value) |
| Input resistance |
2.4MΩ for voltage, 10mΩ or better for current, (50/60 Hz, direct input) |
| Basic accuracy |
±0.1%rdg. ±0.2%f.s. (DC), ±0.1%rdg. ±0.1%f.s. (45Hz to 66Hz)
Note: Provided accuracy of 1 Year, typical value |
| Display refresh rate |
5 times per second |
| Frequency characteristics |
DC, 45Hz to 5kHz |
| Waveform output |
Parameter output representation: voltage, current and active power (3 simultaneous channels), Output voltage: 1V DC f.s. |
Analog output (D/A output) |
Parameter output representation: voltage, current active power and selected 1 item (4 simultaneous channels), Selected 1 item from apparent power, power factor, current integration, active power integration, Output voltage: ±2V DC f.s. |
| Functions |
Rectification method switchable between AC+DC (True RMS), DC (simple average), AC (True RMS), Wave peak measurement, VT or CT ratio settings, Average function |
| Interface |
RS-232C included as standard, GP-IB (Model 3334-01 only) |
| Power consumption |
100V to 240V AC, 50/60Hz, 20VA max. |
| Dimensions and mass |
210 (8.27in) W × 100 (3.94in) H × 245 (9.65in) D mm, 2.5kg (88.2oz) |
| Supplied accessories |
Power cord (1) |
|
|