| Power Measuring Instruments (For use on production lines) |
POWER HiTESTER 3332 |
|
 |
|
 |
 |
|
|
 |
 |
| Can measure very low effective power, for stand-by mode of home use equipment |
 |
- Ultra high-sensitive measurement, for use to measure an effective power of equipment in stand-by mode:
Current 1.0000 mA full-scale, 0.1 µA resolution
- Wide measurement range, up to 50.000 A direct input
|
 |
| Basic specifications |
 |
| Measurement lines |
Single-phase/two-wire |
| Measurement items |
Voltage, Current, Current peak, Active power, Apparent power, Reactive power, Power factor, Phase angle, Frequency, Power integration, Current integration |
| Measurement ranges |
Voltage: 15.000 V to 600.00 V, 6 ranges, Current: 1.0000 mA to 50.000 A, 15 ranges, Power: 15.000 mW to 30.000 kW, 90 ranges, Frequency: 1 Hz to 100 kHz, auto or 2 ranges |
| Integration range |
0 to ±999999 MAh/MWh, (integration time up to 10000 hours) |
| Wave peak measurement |
Current (displays maximum absolute value) |
| Basic accuracy |
±0.1 % rdg. ±0.1 % f.s. (active power, at 45 to 66 Hz) |
| Frequency characteristics |
1 Hz to 100 kHz (1 Hz to 100 kHz at 20 A or less, 1 Hz to 10 kHz at 20 A to 30 A, 50/60 Hz at 30 A to 50 A ) |
| Signal output |
Analog level: Voltage, current, active power, 5 V DC f.s., Waveform monitor: Voltage, current, 1 V rms f.s. |
| Other functions |
Comparator function for 2 items Scaling function (PT/CT/SC), Displays a moving average function, GP-IB/RS-232C interface |
| Sampling rate |
5 times/second |
| Power supply |
100 to 240 V AC, 50/60 Hz |
| Dimensions, mass |
210 mm(8.27 in)W × 100 mm(3.94 in)H × 261 mm(10.28 in)D, 2.5 kg (88.2 oz) |
| Accessories |
Power cord (1), Connector for EXT I/O (1) |
|
|