Japanese Chinese Korea
CLAMP ON AC/DC HiTESTER 3290/3290-10 < Clamp Meters < Products
Clamp Meters(Current Meters -for AC/DC, two ways type-)

CLAMP ON AC/DC HiTESTER
3290/ 3290-10

CE RMS m329010.gif
3290-10+9692
  

3290+9691
OVERVIEW OPTION DOWNLOAD CATALOG  
All the Functions You Need for Measurement at DC or 1Hz and Up
  • Choice of three sensors (Example combinations)
    3290/3290-10 +9691 : Measure up to 100A
    3290/3290-10 +9692 : Measure up to 200A
    3290/3290-10 +9693 : Measure up to 2000A
  • Choice of measurement methods
    DC (for battery measurement)
    AC+DC RMS (for full-/half-wave rectification measurement)
    AC RMS (for current distortion measurement)
    PEAK (for peak value measurement) of inrush current, etc.)
  • Choice of output (Simultaneous output)
    Effective value output, frequency output, waveform output
  • 3290-10 Functions
    Current integral measurement
    (obtain polarity-specific integrated DC values)
    Operating time/duty measurement
Basic specifications (Accuracy guaranteed for 1 year)
  3290 3290-10
Measurement method DC, AC+DC (True RMS), AC (True RMS/ Average rectified) DC, AC+DC (True RMS), AC (True RMS)
Peak value display Absolute value (of waveform peak) Polarities independently (±peak value of waveform, in DC mode)
Separate AC from DC output Separate constituent element of AC, DC None
Output (terminal 1) Current waveform or true RMS value (2 V / setting range), Frequency value (1 VDC / setting range) Current waveform or true RMS value (2 V / setting range)
Output (terminal 2) Rms current/Low-battery warning switch Current integral value (1 VDC / setting range)
Integral current measurement None ±, +, - integral, average value
Rate measurement None Operating time rate, total measurement time
Timer settings None 99 hr, 59 min. (for repeated up to 20 times)
Data storage None Peak, integral value, operating time rate, or other
Basic accuracy at DC current 20.00/ 100.0A (using the 9691 sensor): ±1.3 % rdg. ±0.10 A/ ±0.5 A
20.00/ 200.0A (using the 9692 sensor): ±1.3 % rdg. ±0.10 A/ ±0.5 A
200.0/ 2000A (using the 9693 sensor): ±1.8 % rdg. ±1.0 A/ ±5 A
Basic accuracy at AC current
(45 to 66 Hz)
20.00/ 100.0A (using the 9691 sensor): ±1.3 % rdg. ±0.08 A/ ±0.3 A
20.00/ 200.0A (using the 9692 sensor): ±1.3 % rdg. ±0.08 A/ ±0.3 A
200.0/ 2000A (using the 9693 sensor): ±1.3 % rdg. ±0.8 A/ ±3 A
Voltage measurement None
Frequency measurement 1.00Hz to 1000Hz, Basic accuracy: ±0.3% rdg. ±1dgt.
Crest factor 2.5 or later, or rated peak value of sensor
Frequency characteristic DC, or 1 Hz to 1 kHz (accuracy guaranteed range)
At monitor output: based on the using sensor
Other functions Selectable of responce time, filter (fc=550 Hz)
Power supply LR6 (AA) alkaline batteries × 4, Continuous use : 22 hr, or AC adapter 9445-02/-03 (The 3290-10 only: external power supply possible)
Dimensions, mass 155 mm (6.10 in) W × 98 mm (3.86 in) H × 47 mm (1.85 in) D, 545 g (19.2 oz)
Supplied accessories Carrying strap ×1, LR6 (AA) alkaline batteries × 4