Japanese Chinese Korea
CLAMP ON LEAK HiTESTER 3283 < Clamp Meters < Products
Clamp Meters(Leak Current Meters -for AC only-)

CLAMP ON LEAK HiTESTER 3283

CE
CAT III 300 V
DCL
Max. rated voltage to earth 300V rms insulated wire
RMS
OVERVIEW OPTION DOWNLOAD CATALOG  
Easily monitor leak current fluctuations
  • Can measure a leak current certainly by high-sensitivity of 10 μA resolution (at 10.00 mA range)
  • Indication of 50/60 Hz leak current components by the filtering function
  • Can be used for monitoring the leak current condition in combination with a MEMORY HiCORDER(monitor output)
Basic specifications (Accuracy guaranteed for 1 year, or opening and closing of the clamp sensor 10,000 times)
AC Current 10.00m/ 100.0m/ 1.000/ 10.00/ 200.0 A, 5 ranges, (40Hz to 2kHz, True RMS), Basic accuracy: ±1.0 % rdg. ±5 dgt.
AC Voltage None
Frequency measurement 30.0 Hz to 1000 Hz, 2 ranges, Basic accuracy: ±0.3 % rdg. ±1 dgt.
Crest factor 2.5 and under (1.5 at 200 A range)
Filter function Cut off frequency at filter ON: 180 Hz (-3dB)
Signal Output REC output for current: DC 1 V/f.s. (2 V/f.s. at only 200 A range), response time: less than 200 ms.
MON output for current: AC 1 V/f.s. (2 V/f.s. at only 200 A range), 5 Hz to 15 kHz frequency band width (-3 dB)
Other function Max./ Min./ Average value record, Data hold, Auto-power OFF
Display LCD, max. 2000 dgt., Display refresh rate: 4 times/s at fast, 1 time/3 sec. at slow, and Bar graph display 4 times/s
Power supply 6F22 (Stacked manganese battery) ×1, Continuous use: 40 hours, or AC adapter 9445-02
Core jaw dia. φ 40 mm (1.57 in)
Dimensions, mass 62 mm (2.44 in) W × 225 mm (8.86 in) H × 39 mm (1.54 in) D, 400 g (14.1 oz)
Supplied accessories CARRYING CASE 9399 ×1, Hand strap ×1, Stacked manganese battery (6F22) ×1, Instruction Manual ×1
OVERVIEW OPTION DOWNLOAD CATALOG