| Power Measuring Instruments (Applications to evaluate new energies, inverters and motors) |
MOTOR / HARMONIC HiTESTER 3194 |
|
 |
|
 |
 |
|
|
 |
 |
| Up to 6 input channels for analysis of inverter motors and high-order harmonics! |
 |
- The 3194 provides Analysis of high-order harmonics up to the 3000th order
- This makes it Ideal for analyzing and evaluating the performance of inverter motors and for harmonic analysis of household appliances.
- Additionally, With the optional unit 9603-01 installed, the HiTESTER can directly measure torque and rotation speed, an essential feature for evaluating the performance of inverter motors.
|
 |
| Note: Main unit 3194 cannot operate alone. A factory option input unit the 9600 to the 9604 must be purchased. In the event of unit replacement or extension, the work involved is done at the factory, and unit cost + service fee are charged. |
| Basic specifications (Accuracy guaranteed for 6 months) |
 |
| Measurement lines |
Single-phase 2-wire, single-phase 3-wire, three-phase 3-wire, and three-phase 4-wire systems |
| Number of channels |
Up to 3 channels from input channels 1 to 6, depending on 3194 wiring mode |
| Data outpu |
FD, RS-232C or GP-IB interface, Printer |
| Measurement range |
Fundamental frequency 10 Hz to 4.5 kHz |
|
 |
| Harmonic Waveform Analysis Functions |
 |
| Measurement method |
PLL locked sampling or external clock sampling method |
| A/D resolution |
12-bits |
| Window type |
Rectangular (with gap between windows) |
| Crest factor |
2.5 or less (Voltage and Current) |
| PLL source |
Voltage (U), current (I) or external synchronization of the selected measurement channel |
| External synchronization signal |
Input to a rear panel control terminal on the 3194 main unit
Input level: 1 V to 10 Vrms (for sine waves)
Division function: 1/1 to 1/255 |
| Measurement items |
RMS voltage, RMS current, effective power value, frequency, ±U peak, ±I peak |
| Harmonic waveform measurement items |
Harmonic level, Included percentage of harmonic wave, Phase angle of harmonic wave, Total harmonic distortion |
| Screen disolays |
List, bar-graph, vector and waveform displays |
|
 |
| Other Functions |
 |
| Wiring conversion |
Δ−Y voltage conversion, Y−Δ voltage conversion |
| Sort function |
Sorts according to decreasing order of analysis, displays up to 50th order |
| Averaging function |
Index average with time constant of 1.5 sec |
| Power requirements |
100/120/200/230 V AC, switched automatically , 50/60 Hz, 150 VA max. |
| Dimensions, mass |
430 mm (16.93 in) W × 150 mm (5.91 in) H × 370 mm (14.57 in) D, 13 kg (458.6 oz) (with options) |
| Supplied accessories |
Instruction Manual × 1, Power cord × 1, Connector × 1 |
|
|