| Test function |
Voltage endurance test |
| Testing voltage |
0 to 2.5 kV / 0 to 5.0 kV AC, dual-range configuration |
| Voltage testing method |
Zero-toggle switch |
| Transformer capacity |
500 VA (maximum 30 minutes) |
| Voltage adjustment |
Manually adjusted slidac |
| Waveform |
Mains waveform |
| Frequency |
Mains synchronous |
| Measurement items |
Voltage: 0 to 5 kV AC (Average value rectified, effective value display), Current: 0.01 to 120 mA AC (Average value rectified, effective value display) |
| Judgment function |
Contents: UPPER-FAIL / PASS / LOWER-FAIL (Digital setting window comparator method) |
| Timer section |
Setting range: 0.5 to 999 seconds |
| Interfaces |
EXT I/O, EXT SW, RS-232C |
| Display |
Fluorescent tube display (digital), Analog meter |
| Monitor function |
Output voltage, detection current |
| Power supply |
120 V AC, 50/60 Hz (3158-01), 220 V AC, 50/60 Hz (3158-03), 240 V AC, 50/60 Hz (3158-05) |
| Dimensions, mass |
320 mm(12.60 in)W × 155 mm(6.10 in)H × 263 mm(10.35 in)D
3158-01: 16 kg (564.4 oz), 3158-03/-05: 18 kg (634.9 oz) |
| Accessories |
H.V. TEST LEAD 9615 (high voltage side and return, 1 each), Power cord (1), spare fuse (1) |
|